PARTICLE COUNTER / EN
20
09/15/17 Version No. 00
To connect the ac adapter:
1. Attach the power cord to the end of the ac adapter.
2. Plug the ac adapter into the 12V jack on the meter.
3. Plug the power cord into an ac outlet. The meter will take 2.5 hours to fully
charge.
Note:
1. If the Meter is powered off during charging (Trickle Charge mode), the
battery will require up to 27 hours for a full charge. To charge the battery in
2.5 hours, the meter must be turned on while charging (Fast Charge
mode).
2. When the battery capacity below 25% (
), can not perform particle
count mode and the “Low Battery” mark is displayed.
8-3 Purging the Meter Sensor
Caution
To avoid possible damage to the meter, avoid using the meter in an
excessively dirty or dusty atmosphere.
Excessive particle intake can damage the meter.
To purge the particle counter sensor:
1. Fit the adapter onto the end of the zero count filter.
Note the arrow on the filter denoting airflow direction.
2. Connect the adapter directly onto the end of the isokinetic probe.
3. Operate the unit for approximately 15 minutes.
4. On the average, not more than one particle greater than 0.3
µ
m in five minutes
or not more than one particle per 0.5 cubic foot should be apparent.
Once the unit is purged, resume normal use. Remove the zero count filter and
adapter from the isokinetic prob.