Galaxy S+ CO
2
Incubator
User Guide
#
NOTE:
If the incubator is to be used humidified, the normal RH level is
≈
95%, any
mains powered equipment or device must be designed for use in a humid
atmosphere. Consult the manufacturer of the equipment to ensure that it is
suitable.
To minimise the thermal gradient within the chamber the device should be placed on
the bottom shelf. The heat dissipated by the equipment should be as low as possible
to minimize thermal disturbance to the chamber. Some shakers, stirrers etc. have
been specifically designed to give a low heat output for use in incubators. The
maximum heat that can be dissipated within the chamber is variable, dependant on
the difference between the ambient and programmed temperatures. If the heat
dissipated is too great, the chamber will tend to overheat. If this happens the control
parameters can be adjusted to compensate.
#
NOTE:
With an ambient temperature of 23ºC (maximum) and a programmed
temperature of 37ºC, the maximum power that can be dissipated inside the
incubator is approximately 40 Watts. This figure will be lower if the incubator is
humidified.
Factory testing can be carried out with specific apparatus to ensure that the control
conditions can be met.
Where too much heat needs to be dissipated (>40W) a Galaxy R+ with Cooling
System is required.