Figure 12: BMS Relay Contact Alarm Socket
See Table 5 below for the designation of the six pins.
Table 5: BMS Relay Contact Alarm Socket Pin Designation
Pin
Designation
1 12VDC
unregulated
2 0V
To power external equipment such as a remote buzzer
or light. (100mA maximum)
3 5VDC
Via 10K
Ω
pull-up resistor, provides logic signal to
directly control an auxiliary control system.
4 Normally
closed
5 Common
6 Normally
open
To access the relay contacts. Contact limits are 3 Amps
@ 24V DC and 3 Amps @ 34 VAC.
The default setting for the alarm system is OFF. To activate the relay using the
incubator keypad:
1.
Press the
↵
and
-
keys simultaneously to enter the Alarm Program Mode.
2.
Press the
+
key until
rEL Ay
is shown in the display.
3.
Press the
↵
key to select the option.
4.
The default setting is
nO
(OFF); press the
+
key to change the display to
yEs
(ON) or press the
-
key to return the display to
nO
(OFF).
5.
Press the
↵
key to accept the selection.
6.
To return to the temperature and CO
2
display screen, press the
*
key twice.
8.2
High Temperature Decontamination
High temperature decontamination is designed to heat the internal chamber to
120°C, maintain that temperature for 4 hours, and then allow the chamber to cool
down to 37
º
C (or the programmed temperature) when normal control takes over.
The incubator must be cleaned, decontaminated chemically, and dried thoroughly
before starting the process. The inner and outer door seals must be cleaned
carefully and all surfaces wiped over with decontaminant (see the instructions in
Section 6,
Cleaning & Disinfecting
).
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6