19
RS Biotech
User Guide
Table 4: Temperature & CO
2
Set point Ranges
Parameter
Available Set point Range
Temperature
10
º
C to 50
º
C
(must be at least 1
º
C above ambient)
CO
2
0.2%
to
20%
To set the Temperature:
1.
Press the Programming (
∗
) key. The temperature display will flash.
2.
Press the Up (
+
) or Down (
-
) key until the desired value appears in the left
display.
3.
Press the Enter (
↵
) key to save the setpoint. The CO
2
display will flash.
4.
Now you can set the CO
2
level.
To set the CO
2
level:
1.
Press the up (
+
) or down (
-
) key until the desired value appears in the right
display.
2.
Press the Enter (
↵
) key to save the setpoint.
If at any time you wish to change the CO
2
level without adjusting the temperature set
point, press the Programming (
∗
) key to work in the CO
2
display, then press the
Enter (
↵
) key.
#
NOTE:
Allow the incubator to stabilise at the set points selected before continuing.
3.5
Programming the Alarm System
3.5.1
Setting High & Low Temperature Alarms
1.
Press the
↵
&
-
keys simultaneously to enter Alarm Program Mode. The
display will show
;
°
C
.
AL
.
2.
Press the
↵
key to display the High Temperature Alarm,
HI37
.
5
. The
factory setting is the setpoint value (37.0
°
C) plus 0.5
°
C.
3.
If you wish to adjust the High Temperature Alarm, use the
+
or
-
key. The
minimum setting, however, is 0.5
°
C from setpoint.
4.
Press the
↵
key to save the setting, whether or not you have changed it.
Now the Low Temperature Alarm is displayed,
LO 36
.
5
. The factory
setting is the setpoint value (37.0
°
C) minus 0.5
°
C.
5.
Use the
+
or
-
key if you wish to adjust the value.
6.
Press the
↵
key to accept the setting, whether or not you have changed
it.
7.
Press the Programming (
∗
) key twice to exit the menu.