5.5.1  Creating reports 

The “Report” button can be found on every measurement screen. After a measurement was performed you 
can tap on it and a sidebar menu is shown. 

 

 

Section

: A report is divided into sections which hold the individual measurements. To add a measurement to 

your report you will have to create a section first. If a section already exists you can choose it here. 

 

Current Graphs - Add

: Add this measurement to the chosen section of your report. 

 

Report - Manage

: Show all the saved measurement graphs and create a final report with them. This function 

can also be accessed in the main menu via the button “Reports”. 

 

 

Summary of Contents for IM-A-FI-1800/B3

Page 1: ...PIM Site Analyzer User Manual...

Page 2: ...PIM Site Analyzer Manual Rosenberger 2017 www rosenberger com pia Page 2 of 54...

Page 3: ...surement Screen 18 5 5 Report generation 20 5 5 1 Creating reports 21 5 6 Updating the PIMAnalyzer Software Device Tablet 23 6 Measurement Settings 24 6 1 Manual Mode 24 6 2 2 Tone 25 6 3 Sweep 26 6 4...

Page 4: ...43 11 2 3 IM B FI 700 B12 17 LTE700 L 43 11 2 4 IM B FI 700 B13 14 LTE700 U 44 11 2 5 IM B FI 700 B28 APT700 44 11 2 6 IM B FI 800 B20 DigiDiv800 44 11 2 7 IM B FI 850 B5 AMPS850 44 11 2 8 IM B FI 900...

Page 5: ...sibility for such use of the product The product is used for its designated purpose if it is used in accordance with its product documentation and within its performance limits see data sheet document...

Page 6: ...s permitted only on sockets with an earthing contact and protective earth connection Intentionally breaking the protective earth connection either in the feed line or in the product itself is not perm...

Page 7: ...ctric shock Any object that is not designed to be placed in the openings of the housing must not be used for this purpose Doing so can cause short circuits inside the product and or electric shocks fi...

Page 8: ...erts authorized by Rosenberger Only original parts can be used for replacing safety parts e g power switches power transformers fuses A safety test must always be performed after safety parts have bee...

Page 9: ...vehicle safely and properly The manufacturer assumes no responsibility for accidents or collisions Never use the product in a moving vehicle if doing so could distract the driver of the vehicle Adequ...

Page 10: ...in a special way e g coolants or engine oils that must be replenished regularly the safety instructions of the manufacturer of the hazardous substances or fuels and the applicable regional waste disp...

Page 11: ...for calibration This way the risk of damage during transport can be minimized At first please check if all parts are contained in your shipment otherwise get in contact with Rosenberger See graphic be...

Page 12: ...llows you to perform PIM detection over optical CPRI or RF measurements The plug and play modular and broadband design from 700 2700 MHz with exchangeable band filter units ensures highest flexibility...

Page 13: ...CPRI BBU IN CPRI RRH OUT Test Port The test port connector is easily exchangeable Connection is one of the most important facts causing PIM product During measurement stable connection is necessary s...

Page 14: ...owed but the filter can be changed while the device is running Be sure to protect the connectors on the back of the filter when handling it Integrated Handle Protected Test Port 1 Unlock the filter by...

Page 15: ...equipment The SFP modules in the PIM Site Analyzer have to be compatible with the Base Band Unit or else malfunction may occur To re move the SFP modules flap out the lever to unlock the module and p...

Page 16: ...with the button in the upper left corner It allows you to switch between all the different measurement types and settings 5 2 Common Settings 5 2 1 WiFi Configuration Please refer to chapter 7 Remote...

Page 17: ...ual Rosenberger 2017 www rosenberger com pia Page 17 of 54 5 3 Device Info The Device info screen shows useful information on your PIM Site Analyzer device e g manufacturing and calibration dates of f...

Page 18: ...ply to all other measurement types too Settings Adjust Settings for current measurement Start Stop Start Stop measurement Marker Set Adjust Markers History Show history of measurements Report Create M...

Page 19: ...ith a measurement comment and date time The time span can be changed in the Common Settings screen Check entries to compare several measurement traces or reload them to add them to report The history...

Page 20: ...ow examples of how reports could be structured Site Fridolfing Center 47 997730N 12 824692E Sector 2 PIM Sweep 45 45 PIM 2 Tone 45 tap on surge arrestor 45 tap on surge arrestor 45 bended at jumper 45...

Page 21: ...tion A report is divided into sections which hold the individual measurements To add a measurement to your report you will have to create a section first If a section already exists you can choose it...

Page 22: ...itle and create sections e g Sector East 2 nd Floor or Production Lot1 Production Lot2 Measurements which were added are listed with Frequen cy Band Measurement Type Pass Fail and Date in the selected...

Page 23: ...a to download the latest software version 2 Close PIMAnalyzer application and connected Tablet application 3 Run the Setup file PimAlphaSetup x x exe on the PIMAnalyzer s Windows system typically from...

Page 24: ...tting for CW carrier 1 Power f1 Power setting for CW carrier 1 On Off f2 Enable or Disable carrier 2 Carrier 2 Frequency f2 Frequency setting for CW carrier 2 Power f2 Power setting for CW carrier 2 O...

Page 25: ...Description Stimulus Frequency f1 Frequency setting for CW carrier 1 Frequency f2 Frequency setting for CW carrier 2 Power f1 Power setting for CW carrier 1 Power f2 Power setting for CW carrier 2 ca...

Page 26: ...requency setting for fixed carrier 2 Step size Frequency step size for swept carrier Power f1 Power setting for carrier 1 Downsweep Frequency f2 Frequency from to setting of swept carrier 2 Frequency...

Page 27: ...reflected back instead of pass ing the mismatched component Group Parameter Description Stimulus Frequency Frequency from to setting of swept carrier Step size Frequency step size for swept carrier Po...

Page 28: ...d value for this filter created by zeroing and can only be done on the device not on the tablet 6 5 1 Zeroing PIM At first PIM needs to be zeroed In the Settings Screen of DTF measurement choose PIM L...

Page 29: ...n the DTF settings screen With that the Zero RL button on the measurement screen disappears 6 5 3 Setting the velocity factor of the signal path In order to get an increased accuracy for distance meas...

Page 30: ...M Power f1 Power setting for carrier 1 Power f2 Power setting for carrier 2 must be equal if unit dBc is used Equal Power Couples Power f2 to Power f1 that only one has to be adjusted Error Limit Sets...

Page 31: ...s Therefore a signal is transmitted from the PIM measuring Port 7 16 or 4 3 10 and the leakage received at the N connector Group Parameter Description Stimulus Frequency Frequency from to setting of s...

Page 32: ...meter Description Sweep Setup Frequency f1 Frequency setting for CW carrier 1 Frequency f2 Frequency setting for CW carrier 2 Frequency fIM Shows resulting IM Frequency from f1 f2 setting Power Power...

Page 33: ...sig nals which must be ignored on PIM measurements Group Parameter Description Measurement Setup Frequency Frequency from to setting of monitoring range in RX range Reference Level Level of reference...

Page 34: ...control measurements remotely e g on top of a tower To be able to work remotely you need to activate the WiFi hotspot functionality of your PIM Site Analyzer Open the Common Settings screen from the m...

Page 35: ...search for devices in the connected network As soon as the device you wish to con nect to appears on the results list tap on it and the tablet will be connected to the chosen device If the tablet is n...

Page 36: ...ressurised air or alcohol soaked cotton swaps Ventilation outlets can be cleaned with pressurized air 8 3 Calibration Repair Note that any repairs or calibration of the device not performed by Rosenbe...

Page 37: ...s using appropriate torque 2 Cleaning connector interfaces with alcohol Q Tip and compressed air 3 Replace components e g feeder connector 9 1 1 PIM acceptance test 1 Disconnect cable from base statio...

Page 38: ...ling length an environmental influence is causing the prob lem e g rusty roof surge arrester 5 Fix the located PIM problem 6 Do an acceptance test described in 9 1 1 9 1 3 PIM testing flowchart Spectr...

Page 39: ...to Device Info Debug Data and generate a debug data zip file 6 Close Tablet App 7 On the Device Go to Device Info Debug Data and generate a debug data zip file 8 Send the files to pia_service rosenber...

Page 40: ...PIM Site Analyzer Manual Rosenberger 2017 www rosenberger com pia Page 40 of 54 CPRI link not possible Incompatible SFPs for BBU equipped Change SFPs to match with BBU...

Page 41: ...min typical use non pulsed CW measurement Frequency Stability 2 5 ppm 11 1 2 PIM Analyzer RF Base Unit IM Order 3 rd 5 th 7 th 9 th 11 th 13 th 15 th 17 th Output Power 26 49 dBm equals 23 46 dBm at T...

Page 42: ...1 00 25 00 dB Distance to VSWR fault Accuracy Resolution Range 0 2 m typ Depends on number of mismatch sources and accuracy of cable velocity factor 0 150 m 11 1 6 Spectrum Analyzer RF Frequency Uplin...

Page 43: ...ds on frequency band Weight 3 4 5 kg depends on frequency band Internal Ports 5 P SMP SLC landing socket RF Port Field replaceable Testport 7 16 DIN 4 3 10 DIN opt 11 2 2 Environmental Operating Tempe...

Page 44: ...Path Range TX1 Range TX2 758 775 MHz 782 803 MHz Receive Path Range 703 748 MHz Residual PIM 125 dBm 168 dBc referred to 2 43 dBm Weight approx 4 kg 11 2 6 IM B FI 800 B20 DigiDiv800 ETSI Band B20 Tra...

Page 45: ...dBm Weight approx 4 kg 11 2 9 IM B FI 1400 B11 21 LTE1400 ETSI Band B11 21 Transmit Path Range TX1 Range TX2 1475 9 1487 MHz 1494 1510 9 MHz Receive Path Range 1427 9 1462 9 MHz Residual PIM 125 dBm...

Page 46: ...MHz Residual PIM 125 dBm 168 dBc referred to 2 43 dBm Weight approx 4 kg 11 2 12 IM B FI 2100 B1 UMTS2100 ETSI Band B1 Transmit Path Range TX1 Range TX2 2110 2130 MHz 2150 2170 MHz Receive Path Range...

Page 47: ...M 07P IM 08P IM 85P IM 09P IM 18P IM 19P IM 21P IM 26P IM 35P IM 07P BB IM 08P BB IM 85P BB IM 09P BB IM 18P BB IM 19P BB IM 21P BB IM 0710 BB IM 1822 BB IM 2526 BB IM 3435 BB IM 0722 BB IM A BU 0727...

Page 48: ...28 760 MHz IM B FI 700 B28 APT 700 28 703 748 MHz 758 806 MHz IM B FI 800 B20 DigDiv 20 832 862 MHz 792 822 MHz IM B FI 850 B5 AMPS 800 5 824 849 MHz 869 894 MHz IM B FI 900 B8 EGSM 900 8 880 915 MHz...

Page 49: ...ent and during transportation Allows easy access to all interfaces of Filters can be changed while is packed in bag Material Nylon UV protected Color Dark light grey Dimensions L W H 500 430 200 Weigh...

Page 50: ...10 Female adapter 60S164 S00N1 7 16 Male to 4 3 10 Male adapter 60K164 S00N1 7 16 Female to 4 3 10 Male adapter PIM Standards 60S110 KxxN1 110dBm standard adapter band specific included with every fil...

Page 51: ...senberger Site Solutions GmbH Mayerhofen 45A 83410 Laufen Germany Phone 49 8684 18 5000 Fax 49 8684 18 1499 siso rosenberger de www rosenberger com siso 14 2 Americas Rosenberger Site Solutions LLC P...

Page 52: ...Ltd No 3 Anxiang Road Block B Tianzhu Airport Industrial Zone Beijing 101300 PR China Phone 86 10 80 48 1995 Fax 86 10 80 48 2438 info rosenbergerap com www rosenbergerap com 14 5 India Rosenberger E...

Page 53: ...PIM Site Analyzer Manual Rosenberger 2017 www rosenberger com pia Page 53 of 54 Status May 2017 Technical modifications and errors excepted Similar images...

Page 54: ...1 83413 Fridolfing P O Box 1260 84526 Tittmoning Germany Phone 49 8684 18 0 info rosenberger de www rosenberger com Certified by ISO TS 16949 DIN EN 9100 ISO 9001 ISO 14001 Order No pA 352613 EIM B A...

Reviews: