April 2000 Rosemount Analytical 245364-U
Model 755A Oxygen Analyzer
Example 2, Selection of a Zero Suppression Recorder Oxygen Range:
Desired oxygen range for recorder output: 90% to 100%.
Required span is 10% oxygen; thus, a gain of 10 is selected with the jumper.
Required zero suppression is 90% oxygen; thus the Zero Suppression Selection
Jumper is placed in the 80% position, and the Zero Suppression Potentiometer
is set for 10% oxygen.
S
ECOND
S
TAGE
A
MPLIFIER
This amplifier is an inverting configuration that provides a signal attenuation of 2X,
thus reducing the 10-volt fullscale input signal to obtain a 5-volt fullscale output. This
output is routed to:
Recorder Output Resistor Network
. It provides a jumper-selectable output of 0 to
10 mV, 0 to 100 mV, 0 to 1V, or 0 to 5VDC for a potentiometric recorder.
Current Output Receptacle J1
. This connector accepts the optional plug-in
current-output board.
Dual Alarm Amplifier Circuit
. This circuit drives the optional 638254 Alarm Relay
Assembly.