U
NPACKING AND
I
NSTALLATION
2-2
April 2000 Rosemount Analytical 245364-U
Model 755A Oxygen Analyzer
CAUTION: ENCLOSURE INTEGRITY
With reference to Installation Drawing 642349, any unused cable conduit
openings must be securely sealed by permanent closures in order to provide
enclosure integrity in compliance with personnel safety and environmental
protection requirements. The plastic closures provided are for shipping
protection only.
Note
For NEMA 3R service, all conduit must be connected through approved fittings.
This instrument was shipped from the factory configured to operate on 115 VAC or
240 VAC, 50/60 Hz electric power. Verify that the power source conforms to the
requirements of the individual instrument, as noted on the name-rating plate.
2.4 ELECTRICAL CONNECTIONS
2.4.1 L
INE
P
OWER
C
ONNECTIONS
Electrical power is supplied to the analyzer via a customer-supplied, three-conductor
cable, type SJT, minimum wire size 18 AWG. Route power cable through conduit and
into appropriate opening in the instrument case (see Installation Drawing 642349).
Connect power leads to HOT, NEUT, AND GND terminals on TB1, see Figure 2-1.
Connect analyzer to power source via an external fuse, in accordance with local
codes.
Note
Do not draw power for associated equipment from the analyzer power cable.
2.4.2 R
ECORDER
C
ONNECTIONS
Note
Route recorder cable through a separate conduit, not with power cable or alarm
output cable.
If a recorder, controller, or other output device is used, connector it to the analyzer via
a 24-22 AWG two-conductor shielded cable.
Route the cable through conduit to the analyzer and into the case through the
appropriate opening shown in Installation Drawing 642349. Connect the shield only at
the recorder or computer, if used.
Cable connections and output selection for potentiometric and current actuated