I
NITIAL
S
TARTUP AND
C
ALIBRATION
3-14
April 2000 Rosemount Analytical 245364-U
Model 755A Oxygen Analyzer
3. Check this setting by adjusting the SPAN Control to lower the output below the
setpoint. This will re-energize the relay. Turning R64 above the setpoint will
energize the relay.
S
ETPOINT OF
ALARM 2, LOW
1. With span gas flowing, adjust SPAN Control to read desired alarm setpoint on
display or recorder.
2. Turn setpoint adjustment, R68, clockwise to energy relay.
3. Check this setting by adjusting the SPAN Control to lower the output below the
setpoint. This will re-energize the relay. Turning R68 above the setpoint will
energize the relay.
3.5.2 S
ELECTION OF
D
EADBAND
The desired deadband may be selected with the appropriate trimming potentiometer:
ALARM 1 = R73
ALARM 2 = R78
For any setpoint, deadband is adjustable from 1% of fullscale (counterclockwise limit)
to 20% of fullscale (clockwise limit). Deadband is essentially symmetrical with respect
to setpoint.