I
NITIAL
S
TARTUP AND
C
ALIBRATION
245364-U Rosemount Analytical April 2000
3-5
Model 755A Oxygen Analyzer
3.1.2 R
EADOUT OF
A
PPLIED
Z
ERO
-S
UPPRESSION
V
OLTAGE ON
D
IGITAL
D
ISPLAY
In order to establish the precise zero suppression required for the desired recorder
oxygen range, the actual applied zero suppression voltage may be read on the digital
display.
P
ROCEDURE
:
1. Verify that Recorder Voltage Output Jumper is in the position appropriate to the
recorder: 10 mV, 100 mV, 1 V or 5 VDC.
2. Place the front panel TEST switch in position 4, thus disconnecting the signal input
to permit display of the zero-suppression voltage only.
3. Temporarily place Recorder Oxygen Span Jumper in 100% oxygen (i.e., 1X gain)
position. The digital display will now read the applied zero suppression voltage in
Refer to Table 3-1 for descriptions.
F
IGURE
3-2. M
ODEL
755A I
NTERNAL
A
DJUSTMENTS
L
OCATIONS
R2 R1
R3
R4
R5
R6
R8
R9
CR2
U6
I G O
U3
U2
1
2
3
4
I
G
O
C5
C4
C2
U4
U5
I G O
C3 CR1 C1
U1
J1
16
17
14
13
7
12
9
11
10
8
4
3
5
6
2
1
15
18
19
Door
Control Board