U
NPACKING AND
I
NSTALLATION
245364-U Rosemount Analytical April 2000
2-3
Model 755A Oxygen Analyzer
N
E
U
T
H
O
T
TB1
N
E
U
T
H
O
T
TB1
Jumpers
120 VAC CONFIGURATION
240 VAC CONFIGURATION
+ - +
MA
MV
TB2
NO
COM
NC
RESET
NO
COM
NC
RESET
NO. 1
NO. 2
HOT
GND
N
E
U
T
H
O
T
TB1
COM
Jumper
GND
GND
Power Connections
(see below)
Optional Alarm Kit
mV Recorder
mA Recorder
+
-
+
-
devices are explained in below.
P
OTENTIOMETRIC
O
UTPUT
Insert Recorder Output Selection Jumper, Figure 2-2, in position appropriate to the
desired output; 10 mV, 100 mV, 1V or 5V.
On TB2, Figure 2-1, connect leads of shielded recorder cable to "MV+" AND "COM"
terminals.
Connect free end of output cable to appropriate terminals of recorder or other
potentiometric device.
For device with a span of 0 to 10mV, 0 to 100mV, 0 to 1V, or 0 to 5V, connect cable
directly to input terminals of the device, making sure polarity is correct.
For device with intermediate span, i.e., between the specified values, connect cable to
device via a suitable external voltage divider, as shown in Figure 2-3.
F
IGURE
2-1. E
LECTRICAL
C
ONNECTIONS