3. If still intermittent after replacing
the lead wire, a component may
have failed. Call the repair depart-
ment.
Program option in
use.
Some programs will seem intermit-
tent. This is expected. Refer to the
program.
Option Controls in the Operation
section for a description of the
program option.
Stimulation is
ineffective
Improper electrode
and applicator
placement unknown.
Reposition electrode and applicator.
Contact clinician.
The skin
becomes red
and/or you feel
a stabbing pain.
Using the electrodes
on the same site
every time.
Reposition the electrodes. If at any
time you feel pain or discomfort
stop use immediately.
The electrodes are
not sticking onto the
skin properly.
Ensure the electrode is stuck
securely on the skin.
The electrodes are
dirty.
Clean the electrode pads with a
damp,lint free cloth or replace new
electrode pads. Clean the electrode
belt according to the instructions in
the user manual.
The surface of
the electrode was
scratched.
Replace with new electrodes.
Output current
stops during
therapy
The electrode pads
come off the skin.
Turn off the device and stick the
electrode pad firmly to the skin.
The cable is
disconnected.
Turn off the device and connect
the cable.
The power of the
batteries has been
exhausted.
Replace them with new batteries.