The Rooti RX is intended for use in the electromagnetic environment specified below.
The customer or the user of the Rooti RX should assure that it is used in such an environment.

Emission test                       Compliance                      Electromagnetic environment-guidance

RF emissions
CISPR 11

RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2

Voltage fluctuations
/flicker emissions
IEC 61000-3-3

Group 1

Class B

Class A

The Rooti RX uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.

The Rooti RX is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.

Compliance  

Guidance and manufacturer’s declaration -electromagnetic emissions

The Rooti RX power frequency 
magnetic fields should be at levels 
characteristic of a typical location in a 
typical commercial or hospital 
environment.

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the Rooti RX
requires continued operation during 
power mains interruptions, it is 
recommended that the Rooti RX be 
powered from an uninterruptible power 
supply or a battery.

                      
The Rooti RX is intended for use in the electromagnetic environment specified below.
The customer or the user of the Rooti RX should assure that it is used in such an environment.

Immunity test 

IEC 60601
test level

Compliance level

Electromagnetic
environment-guidance

Electrostatic
discharge(ESD)
IEC 61000-4-2

Electrical fast
transient/burst
IEC 61000-4-4

Surge
IEC 61000-4-5

+ 6 kV contact
+ 8 kV air

+ 1kV line(s) to line(s)
+ 2kV line(s) to earth

+ 2kV for power
supply lines
+ 1kV for input/
output lines

+ 22kV for power
supply lines
Not applicable

+ 1kV differential
mode Not applicable

+ 6 kV contact
+ 8 kV air

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative 
humidity should be at least 30%

Mains power quality should be that of 
a typical commercial or hospital 
environment.

Mains power quality should be that of a 
typical commercial or hospital 
environment.

<5% UT(>95% dip 
in UT) for 0,5 cycle
40% UT(60% dip in 
UT) for 5 cycles
70% UT(30% dip in 
UT) for 25 cycles
<5% UT(>95% dip 
in UT) for 5 s

<5% UT(>95% dip 
in UT) for 0,5 cycle
40% UT(60% dip in 
UT) for 5 cycles
70% UT(30% dip in 
UT) for 25 cycles
<5% UT(>95% dip 
in UT) for 5 s

Voltage Dips, short
interruptions and 
voltage variations 
on power supply
input lines
IEC 61000-4-11

Power frequency 
(50, 60 Hz) 
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Guidance and manufacturer’s declaration-electromagnetic immunity

NOTE  UT is the a.c. mains voltage prior to application of the test level.

                      
The Rooti RX is intended for use in the electromagnetic environment specified below.
The customer or the user of the Rooti RX should assure that it is used in such an environment.

Immunity test  IEC 60601 test level Compliance level

Electromagnetic environment-guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

Portable and mobile RF communications
equipment should be used no closer to any part
of the Rooti RX including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.

    

Recommended separation distance:

d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz

Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in metres (m).
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, should 
be less than the compliance level in each frequency 
range. 
Interference may occur in the vicinity of
equipment marked with the following symbol:

3 Vrms
150 KHz to 80 MHz

3 V/m
80MHz to 2,5 GHz

3 Vrms

3 V/m

Guidance and manufacturer’s declaration-electromagnetic immunity

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 

and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which the Rooti RX is used 

exceeds the applicable RF compliance level above, the Rooti RX should be observed to verify normal operation. 

If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating 

the Rooti RX.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

b

Recommended separation distance between

portable and mobile RF communications equipment and the Rooti RX

The Rooti RX is intended for use in an electromagnetic environment in which radiated RF disturbances are 

controlled. The customer or the user of the Rooti RX can help prevent electromagnetic interference by maintaining 

a minimum distance between portable and mobile RF communications equipment (transmitters) and the Rooti RX 

as recommended below, according to the maximum output power of the communications equipment.

Rated maximum 

output power of 

transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d =1,2√P

80 MHz to 800 MHz

d =1,2√P

800 MHz to 2.5 GHz

d =2,3√P

0,01                            0,12                                      0,12                                        0,23
0,1                              0,38                                      0,38                                        0,73
1                                 1,2                                        1,2                                          2,3
10                               3,8                                        3,8                                          7,3
100                              12                                         12                                           23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in 

metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the 

maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 

and reflection from structures, objects and people.

Summary of Contents for RootRx

Page 1: ...Zhouzi St Neihu Dist Taipei City 114 Taiwan R O C Tel 886 2 2656 0618 E mail service rootilabs com https www rootilabs com Authorised Representative in the European community User Manual V1 0 0598 0560 EC REP Wellkang Ltd www CE marking eu 29 Harley St London W1G 9QR UK ...

Page 2: ...ransmitting single lead EGG data to a qualified receiving station for interpretation by medical professionals The Rooti Rx stores one lead electrocardiography ECG signals and recording data transmission via WiFi To interpret recording data should be skilled at the level of a technician doctor nurse or medical specialist The Rooti Rx comes with the following components Rooti Rx Charging Dock Electr...

Page 3: ... result in minor or moderate injury to the user or patient or damage to the equipment or other property If user feels itchy and allergic after the electrode sticker long time attached to skin please remove the sticker and ask users to CONSULT DOCTOR We advise pregnant women and children to use this product under the supervision of a doctor No modification of this equipment is allowed Be aware of c...

Page 4: ...ing To have a better using experience read the following items Quick RED LED flashing indicates LOW POWER while slow RED LED flashing indicates LEAD OFF When connecting to Rooti Rx users cannot use the internet at the same time If user feels itchy and allergic after the electrode sticker long time attached to skin please remove the sticker and ask users to CONSULT DOCTOR Rooti Rx has to be worked ...

Page 5: ...oad and set up a RootiLink account for your patient 1 After downloaded start RootiLink 2 Sign up a new account for RootiLink set up your personal profile Features ECG EC38 Type 3 Skin Temperature Uses Standard ECG Electrodes Wi Fi data download Rechargeable USB charging Red Green Blue LED Function button IP58 Applications Hospital General Care Skilled Nursing Facilities Home based Remote Patient M...

Page 6: ...till chin up and chest out Attach Rooti Rx on top of patients left chest Make sure the sticker can cover on skin properly without abnormal stretching or twisting Approximately 3cm upper from left nipple ATTENTION a Please use certificated self prepared standard Monitoring Electrodes sold in general pharmacy b During the recording teach patients to short press the button when they feel uncomfortabl...

Page 7: ...Review Record for Your Patients After recording help your patients to download their record in the app Review history ECG data as your further diagnose references 3 Follow the on screen instruction in the app Long press the Rooti Rx function button and activate Wi Fi advertising BLUE LED to recon nect Rooti Rx with RootiLink 4 Tap Download and put the Rooti Rx into Charging Dock Plug into power th...

Page 8: ...ooti Rx it s originally lightening GREEN but it s now no LED signal Rooti Rx is now fully battery charged 0 40 C 20 93 R H 80 0 kPa to 101 3 kPa noncondensing Brand Model Item Measurement Method Rating Power Source Battery Life Operating Environment Storage and transportation Environment Record mode Environment Charging mode and others Main Unit Weight Dimensions for Main Unit Accessory Rooti Root...

Page 9: ...ation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit ...

Page 10: ...1000 4 6 Radiated RF IEC 61000 4 3 Portable and mobile RF communications equipment should be used no closer to any part of the Rooti RX including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80MHz to 800 MHz d 2 3 P 800MHz to 2 5 GHz Where P is the maximum output power rat...