![Ronde & Schwarz R&S FPL1000 User Manual Download Page 681](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989681.webp)
Remote Commands
R&S
®
FPL1000
681
User Manual 1178.3370.02 ─ 09
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 3
<pc>
irrelevant
Parameters:
<NoPowerClasses> 1 to 4
*RST:
1
Example:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Manual operation:
See
"Adding or Removing a Power Class"
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:EXCLusive]
<State>
This command selects the power class used by the measurement if
is set to manual.
Note that:
●
You can only use power classes for which limits are defined.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 3
<pc>
1..n
power class
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation:
See
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
<State>
This command selects the limit check mode for each power class.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 3
Configuring and Performing Measurements