The Spectrum Application (RF Measurements)
R&S
®
FPL1000
206
User Manual 1178.3370.02 ─ 09
Remote command:
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
Detector
Sets the detector for the range.
"Mapping Samples to sweep Points with the Trace Detector"
Remote command:
[SENSe:]LIST:RANGe<ri>:DETector
Reference Level
Sets the reference level for the range.
For details on the reference level, see
Chapter 7.5.1.1, "Reference Level"
Remote command:
RF Attenuation Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see
Chapter 7.5.1.2, "RF Attenuation"
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO
RF Attenuation
Sets the attenuation value for that range.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe
Sweep Points
Sets the number of sweep points for the specified range.
For details on sweep points, see
Chapter 7.6.1.8, "How Much Data is Measured:
Remote command:
[SENSe:]LIST:RANGe<ri>:POINts[:VALue]
Stop After Sweep
This function is not available for the R&S
FPL1000.