
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
107
User Manual 1178.3370.02 ─ 09
Figure 7-6: Measurement of two equally-leveled RF sinusoidal signals with the resolution band-
width which corresponds to the frequency spacing of the signals
Matching generator and R&S
FPL1000 frequencies
The level drop is located exactly in the center of the display only if the generator
frequencies match the frequency display of the R&S
FPL1000 exactly. To achieve
exact matching, the frequencies of the generators and the R&S
FPL1000 must be
synchronized.
5. Set the resolution bandwidth to
100
kHz
.
It is no longer possible to clearly distinguish the two generator signals.
Measurements and Results