Data Management
R&S
®
FPL1000
463
User Manual 1178.3370.02 ─ 09
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
For details on the filename and location, see
Chapter 8.2.2.2, "Storage Location and
Comment
An optional description for the data file. A maximum of 60 characters can be displayed.
Remote command:
File Explorer
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
File Type
Determines whether the global instrument settings with all "Channel"s are stored or
recalled, or the current "Channel" settings only.
Items:
Defines which data and settings are stored or are recalled. Depending on the "File
Type", either channel setups only, or global settings are available. Which items are
available also depends on the installed options (see also
Remote command:
MMEMory:SELect[:ITEM]:HWSettings
MMEMory:SELect[:ITEM]:LINes:ALL
MMEMory:SELect[:ITEM]:TRACe<1...3>[:ACTive]
MMEMory:SELect[:ITEM]:TRANsducer:ALL
Save File
Saves the settings file with the defined filename.
Remote command:
Recall in New Channel / Recall in Current Channel
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific "Channel" only, select "Recall in New Channel" to
activate a new channel setup with the stored settings. Select "Recall in Current Chan-
nel" to replace the current "Channel" settings.