Instrument Construction and Function Description
R&S ZVB
1145.1078.82 3.24
E-4
Putting into operation
Connect the mains cable and turn on at the power on switch. The instrument is now in standby mode.
Insert the BIOS disk in the floppy disk drive.
Turn on the instrument and wait for the first beep. Press the DEL key. The instrument should now
display the setup menu.
•
Select Advanced BIOS Features
•
Enter
•
Select First Boot Device
•
Select Floppy using page up/down key
•
Press F10 key (save)
•
Enter
BIOS has now been programmed.
Do not turn the instrument off when the program is running.
Follow the instructions on the screen.
Select Service Level 2 (see Service Functions).
Check the protocol file for errors:
[ INFO : Error Log ]
Summary of Contents for ZVB 14
Page 4: ......
Page 10: ......
Page 56: ......
Page 58: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 4 ...
Page 66: ......
Page 154: ...Table of Contents Software Update Installation of Options R S ZVB 1145 1078 82 I 4 2 E 4 ...
Page 158: ...5 2 ...
Page 160: ...1145 1078 82 5 2 E 4 Spare Parts List Mechanical Drawings ...
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ...1145 1078 82 5 19 E 4 Block Circuit Diagram ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......