Performance Test
R&S
ZVB
1145.1078.82 1.6
E–4
Checking the Maximum Output Power
Instrument:
Power sensor or
Power meter with power sensor
(see Chapter “Test Equipment“, Item2)
Test setup:
Connect power sensor to port1 (port 2 to port 4)
Power sensor settings:
For measurement frequencies, see Performance Test Report {f
GEN
*}.
R&S
ZVB settings:
– [
Preset
]
– [
Meas
: Wave Quantities: a1 Src Port 1
(a2 Src Port 2 to a4 Src Port 4)]
– [
Sweep
: Sweep Type : CW Mode :
CW Frequency :
{f
GEN
*}
;
Power :
16 dBm
]
– [
Sweep
: Single :
Restart
]
* For measurement frequencies see the Performance Test Report {f
GEN
}.
Measurement:
Read off the level indicated by the power meter.
Max. power ................................................... see Performance Test Report
Summary of Contents for ZVB 14
Page 4: ......
Page 10: ......
Page 56: ......
Page 58: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 4 ...
Page 66: ......
Page 154: ...Table of Contents Software Update Installation of Options R S ZVB 1145 1078 82 I 4 2 E 4 ...
Page 158: ...5 2 ...
Page 160: ...1145 1078 82 5 2 E 4 Spare Parts List Mechanical Drawings ...
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ...1145 1078 82 5 19 E 4 Block Circuit Diagram ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......