![Rohde & Schwarz ZVB 14 Service Manual Download Page 131](http://html.mh-extra.com/html/rohde-and-schwarz/zvb-14/zvb-14_service-manual_1477898131.webp)
R&S ZVB
Instrument Construction and Function Description
1145.1078.82 3.63
E-4
Checking the Temperature Sensors
Select Service Level 2 (see Service Functions).
When the
Hardware Info
softkey is pressed
information about the installed hardware, the
results of temperature measurements and …
are displayed.
Temperature Info
Current Temperature Readings
Component
Sensor
Temperature
Sensor
Temperature
Sensor
Temperature
Motherboard
Near NC:
33.75°C Near SY:
34.75°C Near PS:
38.50°C
Netcon 1
Analog:
40.25°C Digital:
49.75°C
Netcon 2
Analog:
----- Digital:
-----
Reflectometer 1
Generator:
38.50°C Receiver:
42.50°C
Reflectometer 2
Generator:
36.00°C Receiver:
38.75°C
Reflectometer 3
Generator:
38.25°C Receiver:
40.25°C
Reflectometer 4
Generator:
39.25°C Receiver:
41.75°C
Highly elevated temperature values indicate a fan failure.
INFO
Setup
Info
Hardware
Info
Selftest
Error
Log…
Del Error
Log
Summary of Contents for ZVB 14
Page 4: ......
Page 10: ......
Page 56: ......
Page 58: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 4 ...
Page 66: ......
Page 154: ...Table of Contents Software Update Installation of Options R S ZVB 1145 1078 82 I 4 2 E 4 ...
Page 158: ...5 2 ...
Page 160: ...1145 1078 82 5 2 E 4 Spare Parts List Mechanical Drawings ...
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ...1145 1078 82 5 19 E 4 Block Circuit Diagram ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......