![Rohde & Schwarz ZVB 14 Service Manual Download Page 25](http://html.mh-extra.com/html/rohde-and-schwarz/zvb-14/zvb-14_service-manual_1477898025.webp)
R&S
ZVB Performance
Test
1145.1078.82 1.15
E–4
Checking the DC Measurement Inputs
Test equipment:
DC Power Supply NGSM 32/10 (see Chapter “Test Equipment“, Item9)
Multimeter R&S URE3 (see Chapter “Test Equipment“, Item10)
DC cable (see Chapter “Test Equipment“, Item11)
Test setup:
Connect the Power Supply to the Input DC MEAS 1V (DC MEAS 10V) of
the
R&S ZVB using the DC cable.
R&S
®
ZVB settings:
– [
Preset
]
For DC Meas 1 V:
– [
Meas
: More : DC Inputs : DC Meas
±
1 V]
– [
Format
: Real]
– [
Scale
: Scale/Div : .25 x1]
– [
Marker
]
For DC Meas 10 V:
– [
Meas
: More : DC Inputs : DC Meas
±
10 V]
– [
Format
: Real]
– [
Scale
: Scale/Div : 2.5 x1]
– [
Marker
]
Measurement:
Set Power Supply to DC values U
DC
using Multimeter
R&S URE (DC
values see Performance Test Report) and connect it to pos. and neg.
input.
Read off the DC level U
DC ZVB
indicated by the marker.
Calculation:
Deviation = U
DC
–
U
DC ZVB
Summary of Contents for ZVB 14
Page 4: ......
Page 10: ......
Page 56: ......
Page 58: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 4 ...
Page 66: ......
Page 154: ...Table of Contents Software Update Installation of Options R S ZVB 1145 1078 82 I 4 2 E 4 ...
Page 158: ...5 2 ...
Page 160: ...1145 1078 82 5 2 E 4 Spare Parts List Mechanical Drawings ...
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ...1145 1078 82 5 19 E 4 Block Circuit Diagram ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......