Instrument Tour
R&S
®
SMB100A
51
Quick Start Guide 1407.0812.62 ─ 20
[LAN Connector]
Ethernet interface
●
For integrating signal generators in a network
●
Remote control of signal generator
●
Remote access to the signal generator
●
Firmware update
See also:
●
Chapter 4.7, "Connecting to LAN"
●
chapter "Remote Control Basics" in the Operating Manual
[IEC 625/IEEE 488]
IEC-bus (IEEE 488) interface for remote control of the instrument.
See also chapter "Hardware Interfaces" and "Remote Control Basics" in the operating
manual.
Note:
In order to avoid electromagnetic interference (EMI) caused by open lines,
always terminate any connected IEC-bus cable with an instrument or a controller.
[Stereo R/L]
Inputs for analog stereo modulation signals. External modulation sources or the inter-
nal LF generator can be used (stereo modulation is available with option R&S
SMB-
B5).
See also chapter "Stereo Modulation" in the operating manual.
[S/P DIF]
Input for digital stereo signals (stereo modulation is available with option R&S
SMB-
B5).
See also "Stereo Modulation" in the operating manual.
[SIGNAL VALID]
Output of valid signal. This signal marks the valid signal times (valid level and fre-
quency indication). The signal is generated automatically.
[INSTR TRIG]
Input for external trigger for sweeps and list mode.
[PULSE VIDEO]
Output of internal pulse generator signal or external pulse signal fed in via the [PULSE
EXT] connector (video signal).
[PULSE EXT]
Input of external pulse signal or input of external trigger/gate signal for internal pulse
generator.
Rear Panel Tour