Preparing for Use
R&S
®
SMB100A
21
Quick Start Guide 1407.0812.62 ─ 20
b) Mate the connectors along the common axis until the male pin of the connector
of the cable engages with the female socket of the RF connector.
If your instrument is equipped with a test port adapter, see the application note
See
To prevent RF output switch-off
►
NOTICE!
If you set a too high output level without a load connected to the instru-
ment, the reverse power can exceed a limit forcing the R&S
SMB to switch off the
RF output.
Connect a load with sufficient return loss as given in the data sheet.
4.11
Connecting to Ref In/Ref Out
The connector is at the
.
To connect to Ref In/Ref Out
For connection, the R&S
SMB provides BNC connectors.
"To connect to non-screwable connectors (BNC)"
4.12
Switching On or Off
The following table provides an overview of power states, LEDs and power switch posi-
tions.
Table 4-1: Overview of power states
State
LED
Position of power switch
Off
gray
[0]
Standby
yellow
[I]
Ready
green
[I]
To switch on the R&S
SMB
The R&S
SMB is off but connected to power. See
Chapter 4.6, "Connecting to Power"
1. Set the switch on the power supply to position [I].
The LED of the [ON/STANDBY] key is yellow.
Switching On or Off