Preparing for Use
R&S
®
SMB100A
28
Quick Start Guide 1407.0812.62 ─ 20
Changing the default user and security passwords
It is highly recommended to change the default user and security passwords in the
menu "Setup > Security" before connecting the instrument to the network (see section
"Security - Setup-Protection" in chapter "General Instrument Settings" of the Operating
Manual).
4.16.1
Using a Web Browser for Remote Access
The instrument can be remote-accessed via any web browser, as for example the Win-
dows Internet Explorer or Mozilla Firefox.
Alternatively, you can also make use of the LXI browser, as described in
To remote access the instrument via a web browser:
1. Connect the instrument and the remote computer to a LAN, see
2. Instal the Java Runtime Environment JRE on the remote computer.
3. Type the instruments' IP address in the address field of the Web browser on your
PC, e.g.
http://10.111.11.1
The "VNC Authentication" screen appears.
4. Enter the password and select "OK".
The default password is "instrument".
After the connection is established, the current signal generator screen with the
block diagram is displayed and the instrument can be remote-accessed from the
remote computer.
Remote Access via an External Controller