Remote-control commands
R&S
®
SMW-K52/-K116
71
User Manual 1175.6778.02 ─ 19
Parameters:
<Divider>
integer
Range:
2 (R&S
SMW-B10) / 2* (R&S
SMW-B9) to 1024
*RST:
2
Example:
See
Example "Configuring marker settings"
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:OUTPut<ch>:PULSe:FREQuency?
Queries the pulse frequency of the pulsed marker signal
PULSe
.
Return values:
<Frequency>
float
Example:
See
Example "Configuring marker settings"
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:OUTPut<ch>:DELay
<Delay>
Defines the delay between the signal on the marker outputs and the start of the sig-
nals.
Parameters:
<Delay>
float
Range:
0 to 16777215
Increment: 0.001
*RST:
0
Example:
See
Example "Configuring marker settings"
Manual operation:
See
5.6
Clock commands
[:SOURce<hw>]:BB:DVB:CLOCk:MODE
............................................................................71
[:SOURce<hw>]:BB:DVB:CLOCk:SOURce
.........................................................................72
[:SOURce<hw>]:BB:DVB:CLOCk:MODE
<Mode>
Sets the type of externally supplied clock.
Parameters:
<Mode>
SAMP
*RST:
SAMP
Example:
:SOURce1:BB:DVB:CLOCk:SOURce ELCL
:SOURce1:BB:DVB:CLOCk:MODE SAMP
Clock commands