Remote-control commands
R&S
®
SMW-K52/-K116
67
User Manual 1175.6778.02 ─ 19
[:SOURce<hw>]:BB:DVB:TRIGger:SLUNit
<SLunit>
Defines the unit for the entry of the signal sequence length.
Parameters:
<SLunit>
FRAMe | SEQuence
*RST:
SEQuence
Example:
See
Example "Configuring trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:SLENgth
<SLength>
Defines the length of the signal sequence that is output in the
SINGle
trigger mode.
Parameters:
<SLength>
integer
Range:
1 to 7000
*RST:
4
Example:
See
Example "Configuring trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:SOURce
<Source>
Selects the trigger signal source and determines the way the triggering is executed.
Provided are:
●
Internal triggering by a command (
INTernal
)
●
External trigger signal via one of the local or global connectors
–
EGT1|EGT2
: External global trigger
–
EGC1|EGC2
: External global clock
–
ELTRigger
: External local trigger
–
ELCLock
: External local clock
●
Internal triggering by a signal from the other basebands (
INTA|INTB
)
●
In primary-secondary instrument mode, the external baseband synchronization sig-
nal (
BBSY
)
●
OBASeband|BEXTernal|EXTernal
: Setting only
Provided only for backward compatibility with other Rohde
&
Schwarz signal gener-
ators.
The R&S
SMW accepts these values and maps them automatically as follows:
EXTernal = EGT1
,
BEXTernal = EGT2
,
OBASeband = INTA or INTB
(depending on the current baseband)
Parameters:
<Source>
INTB|INTernal|OBASeband|EGT1|EGT2|EGC1|EGC2|ELTRig-
ger|INTA|ELCLock|BEXTernal|EXTernal | BBSY
*RST:
INTernal
Trigger commands