Remote-control commands
R&S
®
SMW-K52/-K116
66
User Manual 1175.6778.02 ─ 19
[:SOURce<hw>]:BB:DVB:TRIGger:OBASeband:DELay
<Delay>
Specifies the trigger delay for triggering by the trigger signal from the other path.
Parameters:
<Delay>
float
Range:
0 to 2147483647
Increment: 0.01
*RST:
0
Example:
See
Example "Configuring trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:OBASeband:DELay
<Delay>
Sets the trigger delay for triggering by the trigger signal from the second path.
Parameters:
<Delay>
float
Range:
0 to 65535
Increment: 0.01
*RST:
0
Default unit: samples
Example:
See
Example "Configuring trigger settings"
[:SOURce<hw>]:BB:DVB:TRIGger:OBASeband:INHibit
<Inhibit>
For triggering via the other path, specifies the duration by which a restart is inhibited.
Parameters:
<Inhibit>
integer
Range:
0 to 67108863
*RST:
0
Default unit: samples
Example:
See
Example "Configuring trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DVB:TRIGger:RMODe?
Queries the signal generation status.
Return values:
<RMode>
STOP | RUN
Example:
See
Example "Configuring trigger settings"
Usage:
Query only
Manual operation:
See
Trigger commands