Connecting the Probe to the DUT
R&S
®
RT
‑
ZM
27
User Manual 1419.3128.02 ─ 04
3.2
R&S
RT-ZMA12 Tip Cable Square Pin
The R&S
RT-ZMA12 tip cable square pin is used to plug onto pins and make a
connection to the DUT that is easy to plug and remove.
The R&S
RT-ZMA12 tip cable square pin is
well suited for creating contact with soldered in
test leads. It can also be used to plug onto pin
strips with a pitch of 1.27 mm (50 mil).
Bandwidth: up to 6 GHz
MultiMode compatible
Length: 15 cm
Distance range: 1.27 mm (50 mil)
Temperature range: -30°C to +80°C
Permanently solder 2 signal (P,N) and ground
wires (GND) to your DUT and effortlessly con-
nect or reconnect your socket head to your
measurement point. In this configuration, you
can measure single-ended, differential and
common mode signals.
However, if you are making only differential
measurements you do not have to connect the
GND lead (GND).
Always keep soldered or connected pins as
short as possible for best performance and
signal integrity.