R&S AM300
Arbitrary and Function Generator
Operating manual, 02/2005
6-137
E-1007.9922.12
6.2.1.6.3
Setting the Anti-Aliasing Filter
Application
You can set different anti-aliasing filters for suppressing interference signals
in your currently applicable waveform:
Auto
Cauer 35 MHz
Bessel 37 MHz
Bessel 75 MHz
External
In the "Auto" setting the Cauer filter (for optimum suppression of image
signals) is automatically switched on for the
Sine
waveform. In the case of
the other waveforms, in which there is optimum settling, a Bessel filter is
automatically switched on.
The "External" setting enables an external filter to be connected to connectors
[15, 29] and/or [16, 15] for the appropriate channel of the R&S AM300, which
is optimized for particular applications.
Selecting the
anti-aliasing filter
1.
In the
submenu
press the
function key
.
A selection field containing specified settings is displayed. The default
setting is "Auto".
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored.
Note
To deliver the output signal to the output complete with all the functions and
parameters that have been set, you need to switch the channel on ( 6-143).
Summary of Contents for AM 300
Page 2: ...02 2005 ...