Remote Control Commands
R&S
®
NRX
409
User Manual 1178.5566.02 ─ 08
●
SCPI Communication Error Queue
SYSTem:ERRor
commands
In the following, the expressions "error number" and "error code" are used synony-
mously.
14.14.9.1
System Event Queue
This system list contains notices, warnings and static errors that happen in setup.
A static error remains in the queue until you solve it. It prevents the execution of nor-
mal measurements. For example, if a power sensor overload occurs, you need to
lower the input power to the power sensor. When you have done that, the static error
disappears.
Warnings and static errors, you can remove from the queue using
.
For further information, see
Chapter 16.2.1, "Interpreting Notifications and Their Num-
Currently, the event type is not included in the description of the event.
.................................................................................................409
..............................................................................................409
SYSTem:SERRor[:ALL]?
Queries all events from the system event queue.
Returns a comma-separated list of unique event numbers.
Usage:
Query only
SYSTem:SERRor:REMove
<num>
Removes a notice or warning from the system event queue. You identify it by its unique
number. To find out the unique number of an event, use
.
Setting parameters:
<num>
Unique number of the event
*RST:
0
Example:
SYST:SERR:REM 3004
Removes the notice that a newer version of the sensor check
source (R&S NRX-B1) is available:
able for sensor check source (R&S NRX-B1)
Example:
SYST:SERR:REM 2007
This number belongs to a static error:
You cannot remove a static error from the queue, thus a "Param-
eter not allowed" error condition occurs. See also
ter 14.14.9.1, "System Event Queue"
System Information and Configuration
www.allice.de