Remote Control Commands
R&S
®
NRX
225
User Manual 1178.5566.02 ─ 08
INITiate<Undef>:ALL[:IMMediate]
Applies to all connected power sensors. See
Suffix:
<Undef>
.
1 to n
No suffix required.
Usage:
Event
14.6
Measurement Settings and Results
The R&S NRX supports up to 4
measurement channels. The trace, pulse analysis, sta-
tistics measurements support 2
traces. Thus, each measurement channel has 2
sepa-
rate measurements, adding up to 8
different measurement results that are configured
using the commands of the
CALCulate<Measurement>
subsystem.
Table 14-1: Assigned measurement panes and traces
Measurement suffix
CALCulate<Measurement>
Assigned measurement panes
WINDow<Window>
Assigned traces
CALC1
to
CALC4
WIND1
to
WIND4
1
CALC5
to
CALC8
WIND1
to
WIND4
2
Measurement pane 1
Trace 1
Trace 2
CALC1
CALC5
Measurement pane 2
Trace 1
Trace 2
CALC2
CALC6
Measurement pane 3
Trace 1
Trace 2
CALC3
CALC7
Measurement pane 4
Trace 1
Trace 2
CALC4
CALC8
For each measurement, you can define two sensor configurations in parallel, a primary
and a secondary sensor configuration.
Further information:
●
Chapter 5.1, "Parallel Measurements"
●
Chapter 14.7, "Calculation Functions"
Measurement Settings and Results