Remote Control Commands
R&S
®
NRX
272
User Manual 1178.5566.02 ─ 08
Further information:
●
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]
............................ 272
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]:AUTO
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
......................................272
[SENSe<Sensor>:]POWer:REFerence
............................................................................. 273
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]
<value>
Sets the reference value for the measured relative values.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<value>
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
on page 56
See
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]:AUTO
<state>
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
ONCE
corresponds to
SET
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<state>
OFF | ONCE
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
<state>
Allows you to relate measured power or a power ratio to a reference value. Whether
the power is measured by one power sensor or whether it is a combined value mea-
sured by two power sensors, is set by:
CALCulate<Measurement>:MATH[:EXPRession]
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH