Remote Control Commands
R&S
®
NRX
302
User Manual 1178.5566.02 ─ 08
<Gate>
1 to 4
Time gate
Parameters:
<value>
Range:
0.0 to 0.1
*RST:
0.0
Default unit: s
Manual operation:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID[:STATe]
<value>
Enables or disables an exclusion interval for the selected gate. This exclusion interval
is called fence. The interval where the fence overlaps with the gate is excluded from
the measurement.
Like a gate, a fence is defined by its start time and its length.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Gate>
1 to 4
Time gate
Parameters:
<value>
*RST:
OFF
Manual operation:
CALCulate<Measurement>[:POWer]:TGATe<Undef>[:AVG]:SELection
<value>
Selects the active gate for the measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Undef>
1 to n
No suffix required.
Parameters:
<value>
Range:
1 to 4
*RST:
1
Manual operation:
14.6.10
Timeslot
In a timeslot measurement, you can use commands that combine several setting com-
mands. They are described in
Chapter 14.4.5, "Timeslot Measurements"
Measurement Settings and Results
www.allice.de