
Remote Control Commands
R&S
®
NRX
371
User Manual 1178.5566.02 ─ 08
Manual operation:
See
[SENSe<Sensor>:]BWIDth:VIDeo:FNUMber
<fnum>
[SENSe<Sensor>:]BANDwidth:VIDeo:FNUMber
<fnum>
Requires the sensor interface for R&S NRT (R&S NRX-B9).
For measuring the peak envelope power, specify the video bandwidth that is used for
measuring the detected RF signal.
Suffix:
<Sensor>
.
101
R&S NRT directional power sensor
Parameters:
<fnum>
Sets the bandwidth of the power sensor as follows:
0 = 4
kHz, 1 = 200
kHz, 2 = maximum bandwidth
Range:
0 to 2
*RST:
0
Manual operation:
See
14.9
Configuring the Test Generator
If the sensor check source (R&S NRX-B1) is installed, you can use it as a power refer-
ence for testing the connected power sensors.
Further Information:
●
●
.................................................................................................371
.................................................................................................371
................................................................................................372
..................................................................................372
OUTPut:SOURce:STATe
<state>
SOURce:OUTPut:STATe
<state>
Requires the sensor check source (R&S NRX-B1)
Enables or disables the signal output.
Parameters:
<state>
*RST:
0
Manual operation:
See
Configuring the Test Generator
www.allice.de
Allice Messtechnik GmbH