Trying Out the Instrument
R&S
®
FSPN
52
Getting Started 1179.4370.02 ─ 01
Integration is done after traces have been smoothed in any way.
1. Enter the "Overview" dialog box and, from there, enter the "Noise Config" dia-
log box.
2. Select the "Integrated Measurements" tab.
3. Select the "Trace" whose data you would like to integrate from the "Show
"dropdown menu.
Tip:
If you select a trace that is currently not displayed in the diagram, no inte-
grated measurement results will be displayed for the corresponding range.
4. Select "Manual" integration ranges.
5. Define the frequency ranges ("Range Start" and "Range Stop") over which
you would like to integrate.
The application calculates the measurement results as defined. Result in the
integrated measurements result table are adjusted accordingly (see
on page 47 for a description on how to view that table).
Figure 5-2: Custom integration range results (frequency range from 3 kHz to 500 kHz,
applied to trace 2)
Note that the integration ranges are displayed graphically in the noise dia-
gram.
Figure 5-3: Custom integration ranges are represented in the diagram as colored lines
For more information on integrated measurement result calculation refer to the
R&S
FSPN user manual.