![Rohde & Schwarz 1173.8709K02 User Manual Download Page 98](http://html1.mh-extra.com/html/rohde-and-schwarz/1173-8709k02/1173-8709k02_user-manual_1477655098.webp)
Remote control basics
R&S
®
NRPxxTWG
98
User Manual 1178.8371.02 ─ 08
9.2.2
Device status register
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Sensor error summary
Sensor error
Sensor error
Sensor error
Sensor error
0
0
Legacy locked
Reference PLL locked
0
0
0
0
0
0
0
Figure 9-2: Device status register
Querying the register:
●
●
Querying the static errors:
●
Table 9-4: Used device status bits and their meaning
Bit no.
Short description
Bit is set if
0
Sum of SERR bits
Sum/combination of SERR bits 1 to 4.
1
SERR measurement not possi-
ble
Static error (SERR) exists. Certain parameter settings could
lead to a situation where subsequent measurements are not
possible.
2
SERR erroneous results
Static error exists. The measurement result is possibly incor-
rect.
3
SERR warning
Static error exists. Status LED of the power sensor is slowly
flashing red.
4
SERR critical
Critical static error exists. Status LED of the power sensor is
fast flashing red.
Status reporting system