Operating concepts
R&S
®
NRPxxTWG
25
User Manual 1178.8371.02 ─ 08
Setup
1
2
3
4
5
6
Figure 5-2: Setup with the R&S
Power Viewer
1 = Signal source
2 = R&S NRPxxTWG power sensor
3 = Host interface connector
4 = R&S
NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed R&S
Power Viewer
1.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
Chapter 3.4, "Connecting to a DUT"
on page 12.
Connect the power sensor to the signal source.
2. Connect the cables as shown in
For a detailed description, refer to
Chapter 3.7.1.1, "Simple USB connection"
Starting a measurement
For a detailed description, refer to the operating manual of the R&S
Power Viewer. The
manual is installed automatically during the installation of the R&S
Power Viewer.
1. Start the R&S
Power Viewer.
2. Execute zeroing.
Note:
Turn off all measurement power signals before zeroing. An active measure-
ment signal during zeroing causes an error.
3. Switch on the test signal of the signal source.
4. Select a measurement.
5. Start the measurement.
5.5
R&S
Power Viewer Mobile
The R&S
Power Viewer Mobile extends the functionality of the R&S
Power Viewer to
Android-based devices, such as a smartphone and tablets.
R&S
Power Viewer Mobile