![Rohde & Schwarz 1173.8709K02 User Manual Download Page 65](http://html1.mh-extra.com/html/rohde-and-schwarz/1173-8709k02/1173-8709k02_user-manual_1477655065.webp)
Remote control commands
R&S
®
NRPxxTWG
65
User Manual 1178.8371.02 ─ 08
Example:
IGAM:MAGN?
Query
1.739179E-02
Response
Usage:
Query only
[SENSe<Sensor>:]IGAMma:PHASe?
Queries the phase angle of the complex input reflection coefficient
Γ
in
. The result is
provided in degrees. Range:
-
180
degrees to +180
degrees.
Example:
IGAM:PHAS?
Query
-1.02
Response
Usage:
Query only
[SENSe<Sensor>:]IGAMma:EUNCertainty?
Queries the expanded (k = 2) uncertainty of the magnitude of the complex input reflec-
tion coefficient
Γ
in
. Following gamma correction, the residual mismatch uncertainty
becomes so small that it is practically negligible.
Example:
IGAM:EUNC?
Query
5.000000E-03
Response
Usage:
Query only
7.8.3.6
Using the S-Parameters program
The S-Parameters program helps loading an S-parameter table into the power sensor.
The S-Parameters program is part of the R&S
NRP Toolkit, see
To start the S-Parameters program
► In the Windows start menu, select "NRP Toolkit" > "S-Parameter Update Multi".
Configuring basic measurement parameters