Power sensor tour
R&S
®
NRPxxTWG
18
User Manual 1178.8371.02 ─ 08
4
Power sensor tour
This chapter provides an overview of the available connectors and LEDs of the power
sensor.
1
2
3
4
Figure 4-1: R&S
NRPxxTWG power sensors (example)
1 = RF connector, see
2 = Status LED, see
Chapter 4.2, "Status information"
3 = Host interface connector, see
4 = Trigger I/O connector, see
Chapter 4.4, "Trigger I/O connector"
4.1
RF connector
The RF connector is used for connecting the power sensor to a device under test
(DUT) or a signal generator. See
Chapter 3.4, "Connecting to a DUT"
For maximum measurement accuracy, tighten the RF connector using a torque wrench
with a nominal torque as specified in the following table.
Table 4-1: R&S
NRPxxTWG RF connector characteristics
Power sensor
Male
connec-
tor
Matching
female con-
nector
Frequency range
Tightening torque*
R&S
NRP75TWG
WR-15
WR-15
50
GHz to 75
GHz
0.58
Nm (5''
lbs)
R&S
NRP90TWG
WR-12
WR-12
60
GHz to 90
GHz
R&S
NRP110TWG
WR-10
WR-10
75
GHz to 110
GHz
* Use the torque wrench for waveguide flanges, R&S
ZCTW, part number 1175.2014.02.
RF connector