Remote control commands
R&S
®
NRPxxTWG
63
User Manual 1178.8371.02 ─ 08
Example:
CORR:SPD:SEL 1
Selects an S-parameter correction data set.
CORR:SPD:STAT ON
Enables the S-parameter correction.
7.8.3.4
S-gamma corrections
Using the complex reflection coefficient, you can determine the power P delivered by
the signal source with considerably greater accuracy.
The coefficient of the signal source
Γ
source
is defined by its magnitude and phase:
●
[SENSe<Sensor>:]SGAMma:MAGNitude
●
The complex reflection coefficient
Γ
sensor
of the sensor, which is also required for the
correction, is prestored in the calibration data memory for many frequencies.
Sensor
Source
Figure 7-2: Correction of interactions between the power sensor and the signal source
If the gamma correction is performed in combination with an S-parameter correction
(
[SENSe<Sensor>:]CORRection:SPDevice:STATe
ON
), the following is consid-
ered:
●
Interaction of the signal source with the S-parameter device
●
Input of the power sensor, depending on the transmission, expressed by the term
s
12
s
21
The interaction between the complex reflection coefficient
Γ
sensor
of the power sensor
and the reflection of port
2 is expressed by
s
22
rection is enabled, this interaction is always considered, regardless whether gamma
correction is performed or not.
Sensor
Source
S
21
S
12
S
22
S
11
Figure 7-3: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands: