STATus:PRESet
Resets the edge detectors and
ENABle
parts of all registers to a defined value.
Usage:
Event
STATus:QUEue[:NEXT]?
Queries the most recent error queue entry and deletes it.
Positive error numbers indicate sensor specific errors. Negative error numbers are
error messages defined by SCPI.
If the error queue is empty, the error number 0, "No error", is returned.
Usage:
Query only
7.13.2
Reading the CONDition part
STATus:DEVice:CONDition?
STATus:OPERation:CALibrating:CONDition?
STATus:OPERation:CONDition?
STATus:OPERation:LLFail:CONDition?
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:SENSe:CONDition?
STATus:OPERation:TRIGger:CONDition?
STATus:OPERation:ULFail:CONDition?
STATus:QUEStionable:CALibration:CONDition?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:POWer:CONDition?
STATus:QUEStionable:WINDow:CONDition?
Usage:
Query only
Using the status register