![Rohde & Schwarz 1145.5850.03 Operating Manual Download Page 79](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-5850-03/1145-5850-03_operating-manual_1477762079.webp)
Operation in Receiver Mode
R&S FSH
1145.5973.12 2.24
E-11
Scanning in the receiver mode:
In the receiver mode, the R&S FSH can scan across a defined number of frequencies and graphically
display the results. It performs a measurement at each frequency for the defined measurement time.
!
Press the SPAN softkey.
!
Press the FREQ SCAN softkey.
The R&S FSH switches to the scan mode and displays the measurement levels at the individual
frequencies as vertical lines.
!
Press the SCAN START softkey.
!
Enter the start frequency for the scan.
!
Press the SCAN STOP softkey.
!
Enter the stop frequency for the scan.
!
Press the SCAN STEP softkey.
!
Enter the step size for the scan.
It is also possible to scan the frequencies of a channel table. This requires activating a channel table as
follows:
!
Press the FREQ softkey.
!
Press the CHANNEL MODE softkey.