Rohde & Schwarz 1145.5850.03 Operating Manual Download Page 213

R&S FSH 

Cable Measurements 

1145.5973.12 4.113 

E-11 

Set the marker on the cable fault to be examined. 

Press the MARKER MODE softkey. 

Using the rotary knob or the cursor keys, select the 
ZOOM ON menu item. 

Confirm with the ENTER key. 

The R&S FSH zooms the display of the cable fault by 
the factor 2. The zoom factor is shown in the entry box 
(here, ZOOM FACTOR: 2). 
 

To zoom in on the fault even more, increase the 
zoom factor by turning the rotary knob or entering 
a higher value. 

The permissible zoom factor depends on the selected 
cable length. The minimum span is 3 m. 
 

The screenshot on the right shows the fault from the 
preceding measurement zoomed by the factor 6. 
Faults mainly at the end of the cable can thus be 
clearly distinguished. 
 

Switch the zoom function off as follows: 
 

Press the MARKER MODE softkey in the marker menu. 

Using the rotary knob or the cursor keys, select the ZOOM OFF menu item. 

Confirm with the ENTER key or by pressing the MARKER MODE softkey again. 

 

Measurement of multiple cable faults using the multimarker function: 

 
If several faults are detected in a cable, the position of each fault can be indicated by a separate marker 
(distance marker) by using the multimarker function. 
 

Press the MARKER MODE softkey. 

Using the rotary knob or the cursor keys, select the MULTIMARKER menu item. 

Confirm with the ENTER key or the MARKER MODE softkey. 

Press the DISTANCE MARKER softkey. 

A menu with six markers opens. 
 

Summary of Contents for 1145.5850.03

Page 1: ...1145 5973 12 11 1 Test and Measurement Division Operating Manual Handheld Spectrum Analyzer R S FSH 1145 5850 03 1145 5850 13 1145 5850 23 1145 5850 06 1145 5850 26...

Page 2: ...1145 5973 12 11 2 Dear Customer R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are trademarks of the owners...

Page 3: ...ching on the Spectrum Analyzer 1 4 Spectrum Analyzer Connectors 1 5 Screen Settings 1 7 Country Specific Settings 1 9 Setting the Date and Time 1 10 Setting the date 1 10 Setting the time 1 10 Chargin...

Page 4: ...ments with the R S FSH Z14 or the R S FSH Z44 2 7 Two Port Transmission Measurements 2 9 Measurement of Return Loss 2 11 Performing Distance To Fault Measurements 2 14 Operation in Receiver Mode 2 20...

Page 5: ...elected 3 2 Entering Measurement Parameters 3 3 Entering values and texts 3 3 Entering units 3 4 Menu Overview 3 5 Frequency entry 3 5 Frequency span 3 5 Level setting 3 5 Bandwidth setting 3 5 Sweep...

Page 6: ...evel 4 8 Entering the display range 4 8 Entering the display unit 4 9 Entering the reference offset 4 9 Entering the input impedance 4 10 Setting the Bandwidths 4 11 Resolution bandwidth 4 11 Video ba...

Page 7: ...l bandwidth 4 48 Displaying the occupied bandwidth 4 48 Changing the span 4 49 Measuring the Carrier to Noise Ratio 4 50 Determining the reference 4 51 DIGITAL Tx mode 4 51 ANALOG TV mode 4 52 CW Tx m...

Page 8: ...the unit for the power readout 4 79 Taking additional attenuation into account 4 80 Two port measurements with the tracking generator 4 82 Measuring the transmission of two ports 4 84 Vector transmis...

Page 9: ...asuring with Transducer Factors 4 122 Unit for measurements with transducers 4 124 Reference level settings for measurements with transducers 4 124 Frequency range of transducer 4 125 Data sets contai...

Page 10: ...dBc 1 Hz 100 kHz from carrier 100 dBc 1 Hz 1 MHz from carrier 120 dBc 1 Hz Sweep time span 0 Hz 1 ms to 100 s span 0 Hz 20 ms to 1000 s min 20 ms 600 MHz Bandwidths Resolution bandwidths 3 dB 1145 585...

Page 11: ...Bm 80 dBm Input related spurious Up to 3 GHz 3 GHz to 6 GHz Signal frequency minus 2 0156 GHz for signal frequencies 2 GHz to 3 2 GHz mixer level 40 dBm carrier offset 1 MHz 70 dBc nominal typ 55 dBc...

Page 12: ...typ 1 5 typ 1 5 Trigger external reference input BNC female selectable Trigger voltage TTL Reference frequency 10 MHz Required level from 50 10 dBm Outputs AF output 3 5 mm mini jack Output impedance...

Page 13: ...8 GHz 8 GHz to 18 GHz 15 C to 35 C 0 C to 50 C 15 C to 35 C 0 C to 50 C 2 3 0 10 dB 4 2 0 18 dB 3 5 0 15 dB 5 0 0 21 dB Zero offset after zeroing 110 pW Dimensions 48 mm x 31 mm x 170 mm connecting c...

Page 14: ...certainty 18 C to 28 C same as for average power plus effect of peak hold circuit Error limits of peak hold circuit for burst signals Duty cycle 0 1 and repetition rate 100 s 20 s repetition rate 100...

Page 15: ...Data Sheet R S FSH3 and R S FSH6 1145 5973 12 TI 6 E 11 Power handling capacity Dimensions 120 mm x 95 mm x 39 mm connecting cable 1 5 m Weight 0 65 kg...

Page 16: ...tainty 200 MHz to 300 MHz 300 MHz to 4 GHz sine signal 18 C to 28 C no zero offset 4 0 of measured value 0 17 dB 3 2 of measured value 0 14 dB Zero offset after zeroing 4 mW Range of typical measure m...

Page 17: ...s plus 1 6 of measured value 0 15 W plus 0 10 W plus 5 of measured value plus 10 of measured value Range of typical measure ment error of peak hold circuit for cdmaOne DAB DVB T CDMA2000 3GPP WCDMA vi...

Page 18: ...wer handling capacity PEAK max 10 ms SWR 3 AVG 10 C to 35 C SWR 1 5 Forward power Frequency 0 2 0 4 0 7 1 2 3 4 GHz 100 200 400 600 800 1000 AVG 10 C to 35 C SWR 3 AVG 35 C to 50 C SWR 3 Dimensions 12...

Page 19: ...to 50 MHz 50 MHz to 3 GHz 3 GHz to 6 GHz typ 20 dB typ 20 dB 12 dB typ 18 dB 16 dB typ 22 dB 16 dB typ 22 dB Return loss at test port corrected 2 MHz to 3 GHz 3 GHz to 6 GHz option R S FSH K2 typ 35...

Page 20: ...and R S FSH6 model 1145 5850 26 Display 301 pixels Maximum resolution distance to fault maximum zoom cable length 1023 pixels Display range Return loss VSWR 10 5 2 1 dB div linear 1 to 2 and 1 to 6 w...

Page 21: ...typ 65 dB typ 80 dB typ 90 dB typ 70 dB typ 85 dB typ 40 dB typ 55 dB typ 35 dB typ 50 dB Reflection measurements only with R S FSH3 model 1145 5850 13 or 1145 5850 23 R S FSH6 model 1145 5850 26 and...

Page 22: ...utes per axis Shock complies with EN 60068 2 27 40 g shock spectrum RFI suppression complies with EMC directive of EU 89 336 EEC and German EMC legislation Immunity to radiated interference Level disp...

Page 23: ...C for the R S FSH R S FSH K1 1157 3458 02 Vector Transmission and Reflection Measurements for the R S FSH R S FSH K2 1157 3387 02 Receiver Mode for the R S FSH R S FSH K3 1157 3429 02 Optional access...

Page 24: ...Spare 50 Load Standard for R S FSH Z2 for VSWR and DTF calibration DC to 3 GHz R S FSH Z31 1145 5780 02 Spare Battery Pack for the R S FSH R S FSH Z32 1145 5796 02 Spare AC Power Supply for the R S FS...

Page 25: ...be changed accordingly 5 If the unit has no power switch for disconnection from the AC supply the plug of the connecting cable is regarded as the disconnecting device In such cases it must be ensured...

Page 26: ...suitably cleaned using a soft lint free dustcloth Never use solvents such as thinners acetone and similar things as they may damage the front panel labeling or plastic parts 12 Any additional safety i...

Page 27: ...sured of receiving a product that is manufactured using the most modern methods available This product was developed manufactured and tested in compliance with our quality management system standards...

Page 28: ...the approximation of the laws of the Member States relating to electrical equipment for use within defined voltage limits 73 23 EEC revised by 93 68 EEC relating to electromagnetic compatibility 89 3...

Page 29: ...onday through Friday from 8 00 until 17 00 If you need assistance outside office hours please leave a message or send us a fax or e mail We will contact you as soon as possible If you wish to receive...

Page 30: ...Buenos Aires Tel 541 14 331 10 67 Fax 541 14 334 51 11 alberto lombardi prec elec com ar Argentina Rohde Schwarz Australia Pty Ltd Sales Support Unit 6 2 8 South Street Rydalmere N S W 2116 Tel 61 2 8...

Page 31: ...Berlin Tel 49 30 34 79 48 0 Fax 49 30 34 79 48 48 info rsv rohde schwarz com Zweigniederlassung B ro Bonn Josef Wirmer Stra e 1 3 D 53123 Bonn Postfach 140264 D 53057 Bonn Tel 49 228 918 90 0 Fax 49 2...

Page 32: ...com jo Jordan Rohde Schwarz Kazakhstan Representative Office Almaty Pl Respubliki 15 480013 Almaty Tel 7 32 72 67 23 54 Fax 7 32 72 67 23 46 rs kazakhstan rsd rohde schwarz com Kazakhstan Excel Enter...

Page 33: ...rabia Gentec Haji Abdullah Alireza Co Ltd P O Box 43054 Riyadh Tel 966 1 293 20 35 Fax 966 1 466 16 57 akanbar gentec com sa Saudi Arabia Rohde Schwarz Representative Office Belgrade Tose Jovanovica 7...

Page 34: ...Mobile Communication P O Box 17466 JAFZ PPU ZG 07 Dubai Tel 971 4 883 71 35 Fax 971 4 883 71 36 Rohde Schwarz Emirates L L C ESNAAD Premisses at Mussafah P O Box 31156 Abu Dhabi Tel 971 2 55 49 411 Fa...

Page 35: ...play RS 232 C optical interface Connector for AC power supply Softkeys Function keys Entry keys Cursor keys Unit keys Generator output N connector RF input N connector Connector for headphones Softkey...

Page 36: ...upper and lower shells The two shells are held together by tape The packaging contains all accessories supplied Undo the tape to unpack the analyzer Upper shell Lower shell CD ROM Headphones Quick St...

Page 37: ...e the optimum viewing angle for the display For use as a desktop fold out the support at the rear so that the instrument can easily be operated from the front and the display can be read easily see Fi...

Page 38: ...he power supply unit Before inserting the power supply unit into the AC power outlet attach the appropriate adapter In vehicles the battery can be charged from the cigarette lighter socket using the R...

Page 39: ...is applied to start a measurement or a 10 MHz reference signal is applied for frequency synchronization The trigger threshold is similar to that of TTL signals The level for the reference signal must...

Page 40: ...for printers with a parallel interface Printers with a USB interface cannot be connected directly to the R S FSH Connector for power sensor The connector has been especially configured for Rohde Schw...

Page 41: ...and confirm by pressing the DISPLAY softkey or the ENTER key again The BACKLIGHT submenu for the lighting level opens The level can be set to HIGH NORMAL and LOW Using the rotary knob or cursor keys...

Page 42: ...the contrast lighting and color settings opens Using the rotary knob or cursor keys select TYPE and confirm with the ENTER key or by pressing the DISPLAY softkey again In the submenu that opens select...

Page 43: ...displayed in a submenu The selected language is highlighted in red Using the rotary knob or cursor keys select the language you want The originally selected language is highlighted in green The red ba...

Page 44: ...to the second field in the date day or month depending on the date format Proceed with the next two fields as with the first After the last data block has been entered the R S FSH verifies the validit...

Page 45: ...five steps until just the battery outline indicates that the battery is flat Battery charge level symbol The battery is charged via the included power supply unit which is connected to the jack on th...

Page 46: ...meters are always to deviate from the default setup for a specific application it is also possible to select a user defined default setup which is then automatically set with the PRESET key This is us...

Page 47: ...set not overwritable or deletable Data set overwritable Status field External Reference External Trigger Switchover The Ext Trig Ext Ref BNC connector on top of the R S FSH can be used either as an i...

Page 48: ...knob or cursor keys select LOW NOISE or LOW DISTORTION Confirm with the ENTER key or the HARDWARE SETUP softkey Using a Preamplifier Only models 1145 5850 03 1145 5850 23 1145 5850 06 and 1145 5850 26...

Page 49: ...el Preamplifier OFF Preamplifier ON Reference level RF attenuation RF attenuation Preamplifier Low Noise Low Distortion Low Noise Low Distortion 25 dBm 0 dB 0 dB 0 dB 0 dB On 24 dBm to 20 dBm 0 dB 0 d...

Page 50: ...0 After you enter your valid PIN the PIN functions can be selected from the selection box When the R S FSH is delivered a new PIN can be activated only if it differs from the factory set PIN Note Befo...

Page 51: ...SH comes with labels reading PIN Code protected If the instrument is protected with a PIN affix one of these labels to the instrument This warns unauthorized users that they cannot operate the R S FSH...

Page 52: ...a parallel interface to the R S FSH using the Serial Parallel Converter R S FSH Z22 thus freeing up the Centronics parallel interface to connect a printer The R S FSH Z22 is powered by a 9 V alkaline...

Page 53: ...sor keys select the printer you want and confirm with the ENTER key or by pressing the GENERAL softkey again The R S FSH displays the selected printer under Printer Type Next set the baud rate for the...

Page 54: ...ENERAL softkey again The R S FSH displays the selected baud rate under SERIAL BAUDRATE in the setup display Enabling Options The R S FSH can be fitted with options e g distance to fault measurements o...

Page 55: ...Installed Options The R S FSH displays the installed options in the Setup menu so you can check them Press the SETUP key Using the rotary knob or the cursor keys scroll the status display downwards Th...

Page 56: ...is used as a signal source e g the Signal Generator R S SML Measurement setup Connect the RF output of the signal generator to the RF input of the R S FSH Signal generator settings Frequency 100 MHz...

Page 57: ...by spectrum analyzers at the top of the measurement diagram is called the reference level REF LEVEL To obtain the best dynamic range from a spectrum analyzer its full level range should be used This m...

Page 58: ...requency from the measurement point frequency and the center frequency and frequency span that have been set The measurement point resolution and consequently the accuracy of the marker frequency read...

Page 59: ...ayed as lines at frequencies that are multiples of 100 MHz To measure the second harmonic ratio set the start and stop frequency as follows Press the FREQ key The softkey menu opens entering the frequ...

Page 60: ...he maximum permissible power for the sensor is never exceeded when measurements are made on high power transmitters Connect the power sensor cable to the R S FSH s power sensor connector and screw it...

Page 61: ...outputs the message Zeroing power sensor please wait When zeroing is over the R S FSH outputs the message Power Sensor Zero OK and switches back to the softkey menu for the power sensor Connect the s...

Page 62: ...terface The cable on the power sensor must be connected and screw fastened to the power sensor connector on the R S FSH The directional power sensor itself has to be inserted between the source and th...

Page 63: ...or 5 CANCEL can be used to cancel zeroing before it begins e g if the signal source cannot be disconnected The R S FSH immediately starts power sensor zeroing While this is being done the R S FSH out...

Page 64: ...d on the R S FSH displays Track Gen Uncal This indicates that tracking generator measurements are uncalibrated Before calibration the span you want should be set because calibration is valid only for...

Page 65: ...FSH displays the magnitude of the transfer function You can read out values with the markers for example The transmission calibration remains valid until the center frequency or the span is changed on...

Page 66: ...s are made This is done with a short and an open at the point were the reflection measurement is to be made If a cable is to be inserted between the DUT and the bridge perform the calibration at the m...

Page 67: ...Using the CONTINUE softkey start the OPEN cali bration While calibration is in progress the R S FSH outputs the message Calibrating OPEN please wait When OPEN calibration is over the R S FSH prompts...

Page 68: ...the Reflection display to indicates possible increase in measurement uncertainty When saving a data set for a scalar reflection measurement in a calibrated state the R S FSH can store the calibration...

Page 69: ...RF and generator port of the R S FSH Z2 R S FSH Z3 to the RF input and generator output of the R S FSH Connect the 1 m test cable supplied with option R S FSH B1 to the bridge test port Note The 1 m c...

Page 70: ...ays the cable used for the measurement in the upper right hand corner of the screen Entering the cable parameters at a specific frequency If cables are used that are not listed in cable models stored...

Page 71: ...meric keys enter the cable length in meters and terminate the entry with the ENTER key or one of the unit keys or Using the rotary knob 1 m steps or the cursor keys 10 m steps adjust the cable length...

Page 72: ...R S FSH displays the return loss produced in the cable under test versus the distance The R S FSH can also list any cable faults It displays the return loss and distance from the measurement plane of...

Page 73: ...ON from the menu Confirm with the ENTER key The entry field for the zoom factor is displayed while the R S FSH simultaneously expands the x axis by a factor of 2 Using the rotary knob or the cursor k...

Page 74: ...in the span for detecting external interferers Press the MEAS MODE softkey Using the rotary knob or cursor keys select SPECTRUM Confirm by pressing the MEAS MODE softkey again or by pressing the ENTE...

Page 75: ...s opens Using the rotary knob or the cursor keys select RECEIVER from the menu and confirm with the ENTER key or the MEASURE softkey The R S FSH activates the receiver mode and measures the level at t...

Page 76: ...R S FSH can also be tuned in channels The channel tables that the R S FSH uses to set channel frequencies are defined either by using the R S FSH View software or by directly entering the first chann...

Page 77: ...cursor keys select NO OF CHANNELS from the menu and confirm with the ENTER key Enter the number of channels and confirm with the ENTER key Press the DEFINE USER TAB softkey Using the rotary knob or th...

Page 78: ...e suitable bandwidth dependent on the set frequency Setting the detector The receiver mode of the R S FSH offers a peak detector average detector RMS detector and quasi peak detector Set the detector...

Page 79: ...ches to the scan mode and displays the measurement levels at the individual frequencies as vertical lines Press the SCAN START softkey Enter the start frequency for the scan Press the SCAN STOP softke...

Page 80: ...of the vision carrier is measured This is common with amplitude modulated TV signals CW Tx In the CW Tx operating mode the power of an unmodulated carrier is measured Manual entry of a reference power...

Page 81: ...Entering the channel bandwidth of the reference channel If the reference measurement is active press the CHANNEL BW softkey Enter the desired value and terminate the entry with the appropriate unit k...

Page 82: ...r measurement is active press the CHANNEL BW softkey Enter the desired value and terminate the entry with the appropriate unit key The R S FSH sets the span in accordance with the selected bandwidth T...

Page 83: ...you can instruct the R S FSH to browse through the list of names of the data sets already stored and display them with the first available free extension Thus the name of the data set recalled for a s...

Page 84: ...h memory space as without it This of course reduces the maximum number of data sets that can be saved In the default state calibration data storage is disabled Press the SETUP key Press the GENERAL so...

Page 85: ...ings and results for each data set are displayed You now have the following options Press the STATUS key to see all the instrument settings in the selected data set When you press the STATUS key again...

Page 86: ...l converter R S FSH Z22 is available Printer with serial interface Connect the printer to the optical interface using the RS 232 C optical interface cable R S FSH Z34 Printer with parallel interface C...

Page 87: ...markers Trace Softkey labelling Battery charge level indicator X axis scale Level axis scale Reference level display Detector display Value entry box Status message Invalid trace symbol Grid Display...

Page 88: ...led The colour of the softkey labelling and its background indicate the status of the softkey function in question Softkey colour Meaning Blue background white labelling Softkey function is turned off...

Page 89: ...letter entry it automatically assigns the letters above the keys to the keys in the alphanumeric keypad The keys have multiple assignments The letter you want is obtained by pressing the key the appro...

Page 90: ...rminate the entry with a unit key Use the unit keys down the right hand side of the alphanumeric keypad These keys have multiple assignments which depend on the unit entry expected by the R S FSH GHz...

Page 91: ...PAN LAST SPAN Level setting AMPT REF RANGE UNIT REF OFFSET RF INPUT 10 dB DIV 5 dB DIV 2 dB DIV 1 dB DIV 0 1 dB DIV LIN 0 100 VSWR 1 REFL COEFF ROH 1 REFL COEFF mROH 1 SMITH CHART 1 2 dBm dBmV dBHV V...

Page 92: ...E SWEEP TRIGGER FREE RUN VIDEO EXTERN EXTERN DELAY Trace setting TRACE TRACE MODE TRACE MEMORY SHOW MEMORY DETECTOR AUTO COUPLED AUTO PEAK MIN PEAK MAX PEAK SAMPLE RMS CLEAR WRITE AVERAGE MIN HOLD MAX...

Page 93: ...CT SELECT USER MOD EXIT DEFINE USER MOD LIST PRINTER THRES HOLD LIST PRINTER EXIT MANUAL SWPTIME Option FSH K2 only STANDARD LEVEL ADJUST PWR UNIT CHANNEL BW POWER DISPLAY USER 3GPP WCDMA CDMAONE CDMA...

Page 94: ...Menu Overview R S FSH 1145 5973 12 3 8 E 11 Power meter menus Transducer menus MEASURE MEAS TRANS DUCER LIMIT LINES DISPLAY LINE TRANSD TRANSD dB EXIT TRD S OFF SELECT TRANSD OFF EXIT LIST PRINTER...

Page 95: ...nus UPPER LIMIT LOWER LIMIT EXIT LIMITS OFF OPTIONS SELECT LIMIT OFF EXIT LIST PRINTER MEASURE MEAS TRANS DUCER LIMIT LINES DISPLAY LINE BEEP MESSAGE VSWR LIMIT EXIT OPTIONS BEEP MESSAGE Smith Chart O...

Page 96: ...MARKERS ON ALL MARKERS OFF DELTA 2 DELTA 3 DELTA 4 DELTA 5 DELTA 6 ALL DELTAS ON ALL DELTAS OFF Distance to fault mode option R S FSH B1 DISTANCE MARKER MARKER MARKER OFFSET SET MARKER MARKER MODE MK...

Page 97: ...E JAPANESE CHINESE KOREAN PRINTER BAUDRATE 1200 Baud 2400 Baud 9600 Baud 19200 Baud 38400 Baud 57600 Baud 115200 Baud PRINTER DESKJET LASERJET EPSON FX POSTSCRIPT PINCODE PINCODE OFF PINCODE ON NEW PI...

Page 98: ...ES DISPLAY LINE ANALYZER RECEIVER TRACKING GEN POWER SENSOR CHANNEL POWER OCCUPIED BW TDMA POWER DISTANCE TO FAULT FREQ REF LEVEL BAND WIDTH MEAS TIME DETECTOR FREQ key Frequency entry FREQ FREQ FREQ...

Page 99: ...START SCAN STOP SCAN STEP Channel entry SPAN FIXED CHANNEL CHANNEL SCAN FREQ START FREQ STOP Channel scan SPAN FIXED CHANNEL CHANNEL SCAN FREQ START FREQ STOP AMPT key AMPT REF RANGE UNIT REF OFFSET...

Page 100: ...ER FREE RUN VIDEO EXTERN EXTERN DELAY Trace key Measurement at a fixed frequency TRACE TRACE MODE TRACE MEMORY SHOW MEMORY DETECTOR TRACE STYLE PEAK AVERAGE RMS QUASI PEAK Scan mode TRACE TRACE MODE T...

Page 101: ...display can be output directly to a printer as measurement documentation At a later date every detail of a measurement can therefore be accurately reproduced Operating sequence Press the STATUS key ab...

Page 102: ...ve you can change the center frequency with the rotary knob or the cursor keys and terminate the entry with the ENTER key You can clear the value entry box from the screen by pressing the CANCEL key T...

Page 103: ...setting the step size is equal to 10 of the span 1 division on the vertical scale With CENTER the step size is equal to center frequency This setting is ideal for measurements on harmonics On each fr...

Page 104: ...R S FSH3 which is greater than 3 GHz or you reach the 3 GHz limit with the rotary knob or the cursor keys the R S FSH3 outputs the message Maximum reached The limit on the R S FSH6 is 6 GHz Working wi...

Page 105: ...the definition of the channel table used either Minimum reached is displayed on the screen when the lowest channel number is reached or Maximum reached is displayed when the highest channel number is...

Page 106: ...MHz kHz or Hz or Change the span with the rotary knob or the cursor keys The span is set immediately after the change is made The value entry box can be cleared from the screen with the CANCEL key Us...

Page 107: ...he RF attenuation but also the preamplifier if switched on is coupled to the reference level setting Preamplifier OFF Preamplifier ON Reference level RF attenuation RF attenuation Preamplifier Low Noi...

Page 108: ...hlighted in red Enter a reference level with the number keys and either terminate the entry with one of the unit keys dBm or dBm for relative measurements or m O n for absolute measurements or press t...

Page 109: ...ing the rotary knob or the cursor keys select the unit you want and confirm your selection by pressing the ENTER key The reference level unit is immediately set on the R S FSH Entering the reference o...

Page 110: ...softkey A submenu opens The two input impedance options 50 and 75 are displayed Select the input impedance you want using the rotary knob or the cursor keys and confirm your selection by pressing the...

Page 111: ...has an effect on the noise displayed by the spectrum analyzer If the bandwidth is small the noise displayed drops If the bandwidth is reduced or increased by a factor of 3 the noise displayed drops o...

Page 112: ...Enter the resolution bandwidth you want using the number keys and terminate the entry with the appropriate unit MHz kHz or Hz or Change the resolution bandwidth to the value you want using the rotary...

Page 113: ...ed the high frequency noise components above the cutoff frequency of the video filter will be rejected The smaller the video bandwidth the smaller the noise amplitude at the video filter output Theref...

Page 114: ...he resolution bandwidth RBW a small red circle is placed before the video bandwidth readout in the top right hand corner of the screen Enter the video bandwidth you want with the number keys and termi...

Page 115: ...red circle on the right hand side of the measurement diagram The R S FSH requires a minimum sweep time of 20 ms per 600 MHz of span If a larger span is set the R S FSH automatically adapts the minimu...

Page 116: ...the CONT SWEEP softkey The R S FSH now sweeps continuously again Trigger To respond to events the R S FSH has a variety of trigger functions The trigger can either be external or generated internally...

Page 117: ...e of the reference level 100 means that the trigger level equals the reference level 50 that the trigger level is in the middle of the y axis on the measurement diagram default setting The position of...

Page 118: ...E 11 The trigger delay resolution depends on the delay selected The table below lists the values Trigger delay DELAY Resolution 0 to 1 ms 10 Os 1 ms to 10 ms 100 Os 10 ms to 100 ms 1 ms 100 ms to 1 s...

Page 119: ...ith the trace pixels from the previous setting for example if the span is changed Intermittent signals in the spectrum or the maximum of fluctuating signals are easy to find with MAX HOLD MIN HOLD The...

Page 120: ...ector is pixel oriented i e it determines how the level at each pixel will be displayed The R S FSH always measures the whole spectrum However the trace only has 301 pixels in the x direction for disp...

Page 121: ...the greater the sweep time If you are making noise measurements for example the trace will be highly stable if a long sweep time is selected However the bandwidth occupied by the signal to be measured...

Page 122: ...the ENTER key or the DETECTOR softkey The R S FSH indicates the detector that has been selected in the top left hand corner of the screen Detect Auto Pk in Fig above If AUTO COUPLED is switched on and...

Page 123: ...y Therefore when the memory trace is recalled it will not be adapted to any modifications of the reference level or span that may have been made in the meantime When a stored data set is called the R...

Page 124: ...the main marker in other words the delta marker frequency is the frequency difference between the frequency at the point marked by the main marker and the frequency at the point marked by the delta m...

Page 125: ...er position PEAK This function places the marker or the delta marker on the highest value of the trace The function acts on the active marker whose associated softkey labelling is highlighted in red N...

Page 126: ...onfirm your choice with the SET MARKER softkey or the ENTER key The R S FSH opens a submenu for setting the start and stop frequencies for the marker search range For entering the start of the search...

Page 127: ...nals in a trace the R S FSH has the multimarker function Up to six different markers are available in the multimarker mode Marker 1 measures in absolute units Markers 2 to 6 can measure in absolute un...

Page 128: ...the marker or delta marker near the position you want The step width here is 10 of the X axis Then use the rotary knob to fine tune the marker or delta marker to the signal The step width corresponds...

Page 129: ...a markers If you press the MKR LIST VIEW softkey again or any other softkey in the marker menu the R S FSH closes the marker table Deactivating markers In multimarker mode markers can be deactivated o...

Page 130: ...ternatively you can exit the multimarker mode by deactivating all markers MARKER key MARKER softkey ALL MARKERS OFF menu item or with PRESET Marker functions Apart from displaying the level and freque...

Page 131: ...301 frequency coordinates The frequency resolution is therefore relatively coarse especially if a large span is set To circumvent this problem the R S FSH s internal frequency counter can be used Whe...

Page 132: ...ically turns on the marker and positions it on the trace maximum Using the cursor keys or the rotary knob select the demodulation mode AM or FM you want and confirm your selection with the ENTER key N...

Page 133: ...s Terminate the entry with the ENTER key The softkey label DISPLAY LINE is printed on a green background and the entry box is cleared In contrast to the markers the position of the displayed line is p...

Page 134: ...olution bandwidth that is small in comparison with the channel bandwidth The measured values on the trace are then integrated to give the total power The R S FSH takes into account the selected displa...

Page 135: ...key or the STANDARD softkey The R S FSH sets the selected standard The optimal span resolution bandwidth video bandwidth sweep time and detector for the standard are selected automatically If USER is...

Page 136: ...g used by the R S FSH in the USER setting is immediately clear The name entered as the USER standard also appears on the screen thus making it possible to document the setting along with the measureme...

Page 137: ...l within the measurement diagram To prevent the R S FSH from being overdriven the signal can be measured at the largest resolution bandwidth possible using the peak detector If this setting is selecte...

Page 138: ...is 833 Hz at a span of 1 kHz Changing the span The span set by the R S FSH yields extremely precise measurement results However signals in the environment of the measurement channel are no longer dete...

Page 139: ...ursor keys or the rotary knob and confirm with the ENTER key The power display will switch from Max Power to Power Power display The R S FSH displays the power at the bottom of the measurement diagram...

Page 140: ...different units for power output The basic unit is dBm Press the PWR UNIT softkey The R S FSH opens the submenu with the units dBm dBmV and dBOV Using the rotary knob or the cursor keys select the re...

Page 141: ...le signal lies within the selected resolution bandwidth If the resolution bandwidth is too narrow the displayed power will be lower than the actual power Press the MEAS key Press the MEASURE softkey T...

Page 142: ...dard also appears on the screen thus making it possible to document the setting along with the measurement Press the STANDARD softkey A table with the available standards opens Select Rename USER with...

Page 143: ...n the greatest possible dynamic range for burst signals the lowest reference level possible must be set If this is not done the R S FSH will be overdriven by the measurement signal if its maximum leve...

Page 144: ...the trigger the R S FSH will trigger on the rising edge of the burst Should this not be the case the trigger level must be adjusted so that the R S FSH is triggered by the burst edge Otherwise no mea...

Page 145: ...entered the R S FSH automatically selects the measurement parameters so that an optimal result is attained Operation Press the MEAS key Press the MEASURE softkey The R S FSH will open the measurement...

Page 146: ...nges to the setting so that it is again available when USER is called again The following should be noted when changes to the settings are made The span is always coupled to the channel bandwidth CHAN...

Page 147: ...corner of the screen after the USER standard is selected Using the R S FSH View control software additional standards can be generated and permanently loaded into the R S FSH You can also delete the f...

Page 148: ...entry with the appropriate unit or Change the channel bandwidth with the rotary knob or the cursor keys and terminate the entry with the ENTER key or the CHANNEL BW softkey The R S FSH automatically...

Page 149: ...pied bandwidth is set MANUAL SPAN entry is active for immediate entry of another span Using the numeric keys enter a new span and terminate the entry with the appropriate unit or Change the span with...

Page 150: ...d is common with amplitude modulated TV signals CW Tx In the CW Tx operating mode the power of an unmodulated carrier is measured Manual entry of a reference power or reference level The R S FSH also...

Page 151: ...following The span is preset to 1 2 times the reference channel bandwidth The span is always coupled to the channel bandwidth When changes are made the R S FSH automatically sets the appropriate span...

Page 152: ...menu for selecting the reference measurement opens Using the rotary knob or the cursor keys select ANALOG TV and confirm your choice with the ENTER key or the F1 softkey The R S FSH starts the ANALOG...

Page 153: ...ng mode the count marker is automatically activated After each sweep the marker is set to the highest power in the reference channel Marker power and marker frequency are indicated at the top edge of...

Page 154: ...e noise channel measurement has been activated the NOISE MEASURE softkey is highlighted in green If DIGITAL Tx is set the R S FSH measures the power in accordance with the selected channel bandwidth M...

Page 155: ...in comparison with the signal bandwidth the R S FSH may be overdriven although the trace is within the measurement diagram To prevent the R S FSH from being overdriven the signal can be measured at t...

Page 156: ...y or the F2 softkey The C N reference is displayed in the selected unit Measuring the noise channel power and calculating the carrier power noise power The noise channel power is measured in an unoccu...

Page 157: ...hat are more stable Press the NOISE MEASURE softkey to start the measurement For calculation purposes the R S FSH uses the reference value that was last measured or manually set During the noise chann...

Page 158: ...3 models 1145 5850 03 and 1145 5850 13 If you attempt to enter a smaller channel bandwidth the R S FSH will automatically set 8 33 kHz and output the message Limit exceeded With the R S FSH3 model 114...

Page 159: ...red C N or C No value C N measurement result display If the noise channel measurement is active the R S FSH inserts the C N measurement result at the bottom of the measurement diagram Usually this doe...

Page 160: ...mum span for the channel power measurement is set MANUAL SPAN is activated to allow immediate entry of another span Using the numeric keys enter a new span and terminate the entry with the appropriate...

Page 161: ...s on specified discrete frequencies using the selected measurement time per frequency Activating the receiver mode Press the MEAS key Press the MEASURE softkey The measurement function menu will open...

Page 162: ...y knob or the cursor keys or enter a new receiving frequency via the numeric keypad The frequency is displayed immediately after the entry Setting the frequency step size The step size for tuning the...

Page 163: ...les available in the instrument that were loaded with the R S FSH View software are displayed If no channel table is loaded No bands available is displayed Select the desired channel table using the r...

Page 164: ...ELS from the menu and confirm with the ENTER key Enter the number of channels and confirm the entry with the ENTER key Press the DEFINE USER TAB softkey Select CHANNEL SPACING from the menu and confir...

Page 165: ...mediately activates entry of the resolution bandwidth the MANUAL RES BW softkey is highlighted in red Change the current bandwidth with the cursor keys or the rotary knob or enter a new bandwidth via...

Page 166: ...bandwidth The quasi peak evaluation for the C D band 30 MHz to 1000 MHz is coupled to the 120 kHz bandwidth The detector is selected either in the main menu of the receiver mode or by using the TRACE...

Page 167: ...of the scan via the numeric keypad or change the start frequency using the rotary knob or the cursor keys Press the STOP SCAN softkey Enter the stop frequency of the scan via the numeric keypad or ch...

Page 168: ...lighted in red in the table of channel tables The screenshot above shows a measurement using a channel table with frequency gaps Different segments that are not interconnected are specified using the...

Page 169: ...or connector and screw into position The DUT is connected to the N connector on the power sensor Power Sensor R S FSH Z1 or R S FSH Z18 RF Connector Power Sensor Connector The continuous power applied...

Page 170: ...e R S FSH outputs the following error messages Message Cause Remedy Error in zeroing signal at sensor A signal was present at the power sensor when zeroing was performed Unscrew the power sensor from...

Page 171: ...want to perform measurements on another known signal the power sensor can be told what the center frequency is via the frequency entry mode FREQ softkey Press the FREQ softkey The frequency value ent...

Page 172: ...puts the message Power sensor zero OK and switches back to the softkey menu for the power sensor Selecting the unit for the power readout The R S FSH can display measured power in relative units dBm o...

Page 173: ...a short measurement time to produce a stabile accurate result In this case the FAST operating mode is recommended to obtain a high repetition rate for the measurement When the NORMAL setting is select...

Page 174: ...FSET softkey The value entry box for the reference offset opens Using the rotary knob the cursor keys or the number keys enter the offset you want and confirm the entry with the ENTER key The offset i...

Page 175: ...orresponding to the direction of forward power The directional power sensors are controlled and powered via a special serial interface The sensor cable is to be connected and screwed to the POWER SENS...

Page 176: ...atus field and after a few seconds displays the type of the sensor connected R S FSH Z44 as well as the measured power In the event of incorrect operation or sensor malfunction the R S FSH outputs the...

Page 177: ...nter frequency to the power sensor If a signal at another frequency is to be measured the new center frequency can be transferred to the power sensor by entering the frequency FREQ softkey Press the F...

Page 178: ...er sensor Setting the power measurement weighting For forward power display the R S FSH provides both average power and peak envelope power Use the POWER DISPLAY softkey in the Power Sensor menu to sw...

Page 179: ...g the units for forward power and reflected power display opens Using the rotary knob or the cursor keys select the parameter for which a unit is to be entered Confirm the parameter with the ENTER key...

Page 180: ...r a number of common transmission standards Press the STANDARD softkey A menu with the selectable standards opens Select the desired standard using the rotary knob or the cursor keys Confirm with the...

Page 181: ...aximum input level of the R S FSH Z14 or R S FSH Z44 a directional coupler or an attenuator has to be connected ahead of the power sensor In such cases the coupling attenuation of the directional coup...

Page 182: ...ion and reflection measurements the magnitudes are corrected Vector calibration methods and measurements option R S FSH K2 are possible for increasing the dynamic range and the measurement accuracy Th...

Page 183: ...or the frequency range are selected with the appropriate keys as in the spectrum analyzer mode When the MEAS key is pressed the softkey menu for the tracking generator is displayed Before calibration...

Page 184: ...put and the DUT s output to the RF input of the R S FSH The R S FSH measures the magnitude of the DUT s transmission The operating sequence is explained below using a transmission measurement on a SAW...

Page 185: ...oftkey F1 or F2 CONTINUE to start calibration To abort calibration press the fourth or fifth softkey CANCEL During calibration the R S FSH outputs the message Calibrating THROUGH please wait Calibrati...

Page 186: ...en such a data set is recalled measurements can be performed without prior calibration providing the instrument s temperature does not deviate more than 5 C from its temperature when the data set was...

Page 187: ...The trace offset function is useful if it is necessary to compensate for a fixed loss or gain during the measurement Vector transmission measurement only with option R S FSH K2 With vector measuremen...

Page 188: ...the tracking generator output without the DUT Press the F1 or F2 CONTINUE softkey to start calibration By pressing the F4 or F5 CANCEL softkey calibration can be aborted During calibration the R S FS...

Page 189: ...To eliminate having to repeatedly mount and dismount the VSWR bridge the VSWR Bridge R S FSH Z3 10 MHz to 6 GHz includes a switch that bypasses the VSWR bridge and simultaneously switches the trackin...

Page 190: ...of the VSWR bridge is incorporated into the measurement This corrective step requires that the control cable for automatically detecting the bridge be connected to the R S FSH Since the value is merel...

Page 191: ...s setting may be useful if you only work with the R S FSH Z3 connected and do not want to waste time on automatic detection Supplying DC voltage to active DUTs By using the VSWR Bridge R S FSH Z3 DC v...

Page 192: ...Connect an open to the measurement port of the bridge By pressing the first or the second softkey CONTINUE start the OPEN calibration While calibration is in progress the R S FSH outputs the message...

Page 193: ...e measurement port of the VSWR bridge The R S FSH displays the return loss of the DUT Entering the display unit Press the AMPT key Press the RANGE softkey The submenu for selecting the display ranges...

Page 194: ...weep time or measurement range has no effect on measurement accuracy This means they can be changed after calibration without any reduction in accuracy When a data set for scalar reflection measuremen...

Page 195: ...or the MEAS MODE softkey Calibration Before the R S FSH is calibrated the desired center frequency and span must be set If they are set later the calibration values are lost and the calibration must b...

Page 196: ...measurement uncertainty must be anticipated The R S FSH retains the calibration data but displays a red dot before the Refl vect display in the top right hand corner of the screen to indicate a possi...

Page 197: ...the Smith chart Press the MARKER key The R S FSH activates the marker menu and a marker Using the rotary knob cursor keys or numeric entry move the marker on the complex impedance trace The numeric ou...

Page 198: ...e displayed VSWR limit value to the desired value by using the rotary knob or enter a new limit value by using the numeric keys Press the ENTER key to complete the entry To deactivate the VSWR limit v...

Page 199: ...e the size of the zoom window or the zoom factor Press the ZOOM FACTOR softkey Select the desired zoom factor 2 4 or 8 by using the rotary knob or cursor keys or by entering a numeric value Confirm th...

Page 200: ...ement on an antenna for example it is useful to switch to the spectrum display To eliminate having to dismount the VSWR bridge each time a typical value for the insertion loss of the VSWR bridge is in...

Page 201: ...by using the rotary knob or the cursor keys and confirm the entry with the HARDWARE SETUP softkey or the ENTER key If you select AUTO DETECT the VSWR bridge is detected automatically as soon as the co...

Page 202: ...ftkey Set the required frequency range on the R S FSH Switch on vector measurement MEAS key MEAS MODE softkey VECTOR menu item Calibrate the R S FSH MEAS key RELECT CAL softkey Connect the cable to be...

Page 203: ...eflected at a fault relative to the generator signal there is either reinforcement or cancellation Because of this effect there is ripple on the received sum signal in the frequency domain The R S FSH...

Page 204: ...the cable Cable selection Frequency dependent cable models can be generated with the supplied R S FSH View Windows software package and loaded into the R S FSH The procedure is described in the R S FS...

Page 205: ...at a user defined cable model has been selected The cable model can be checked or modified using the DEFINE USER MOD softkey Press the DEFINE USER MOD softkey A menu for entering the frequency the vel...

Page 206: ...iberately entering a cable length that is too short is a good way of increasing distance to fault accuracy for a fault that is near to the measurement plane If the entered cable length is greater than...

Page 207: ...e span and then the center frequency This prevents a message from being output stating that the desired center frequency cannot be set for the span currently being used for the distance to fault measu...

Page 208: ...cted as an OPEN is not defined as precisely as a SHORT Note The reference plane must be the output of the 1 m measurement cable i e the measurement cable may not be dispensed with If the output of the...

Page 209: ...valid however the instrument temperature must not change by more than 5 C after calibration If the temperature changes by more than 5 C a red circle is placed ahead of DTF CAL to indicate that there...

Page 210: ...l other units a window for selecting the display range scaling opens You can select a display range by using the cursor keys or the rotary knob Confirm the selection by pressing the ENTER key or the R...

Page 211: ...ey Locating cable faults by means of the marker function The distance to cable faults or the distance between any two faults can also be read off with the marker Press the MARKER key The R S FSH opens...

Page 212: ...s step 10 of the span As is the case with spectrum analysis the R S FSH provides functions to automatically position the marker or the marker offset on the trace These can all be accessed by pressing...

Page 213: ...m the preceding measurement zoomed by the factor 6 Faults mainly at the end of the cable can thus be clearly distinguished Switch the zoom function off as follows Press the MARKER MODE softkey in the...

Page 214: ...n be defined for the distance to fault measurement If a marker offset is defined all distance values output by the R S FSH are referenced to the position of the offset marker Automatic marker position...

Page 215: ...indicate that the R S FSH is measuring return loss DTF refl cal is displayed in the top right hand corner of the screen Further information Setting the span If automatic setting AUTO SPAN is used the...

Page 216: ...re N number of measurement points SPAN manually set span AUTOSPAN span used by the R S FSH in automatic setting of the span In the above measurement example the resolution is thus 202 points for a cab...

Page 217: ...orrection values from calibration It then calculates the IFFT result into length units from the cable parameters light velocity and frequency range In addition the R S FSH considers the attenuation of...

Page 218: ...mask on the screen remains unchanged The points on the y axis are always dB values If the scale on the y axis is linear unit V or W the R S FSH automatically switches to the respective dB unit after a...

Page 219: ...dist Distance DTF distance to fault measurements It is also indicated whether the limit lines are assigned to absolute frequency time or distance values abs or whether they are specified relative the...

Page 220: ...alues are within specified limits PASS is displayed at the top in the center of the diagram FAIL is indicated even if only a single measured value pixel of the trace exceeds a limit value As long as a...

Page 221: ...splayed span a check is not performed outside the definition range Data sets containing limit lines The R S FSH stores data sets together with any limit lines that may have been active for the measure...

Page 222: ...s in this chapter Interpolation between the values is performed with the aid of a modified spline algorithm Even if only relatively few values such as maxima minima and turning points are available th...

Page 223: ...With EXIT the transducer menu can be quit with TRD S OFF all transducer factors can be switched off Press the TRANSD softkey The R S FSH displays a list of transducer factors available in the unit Th...

Page 224: ...t for measurements with transducers If the unit of the transducer is dB the units dBm dBmV or dBOV remain unchanged The linear units Volt and Watt are not permissible They are deactivated in the units...

Page 225: ...cer values outside the defined range to be zero Data sets containing transducer factors The R S FSH stores data sets together with any transducer factors that may have been active for the measurement...

Page 226: ...g transducer factors The transducer factors are antenna specific and are supplied together with the TS EMF antenna Transducers are loaded into the R S FSH by means of the R S FSH View control software...

Page 227: ...o the measurement result EXIT exits the transducer menu and TRD S OFF deactivate all transducer factors Press the TRANSD softkey The R S FSH will open the menu for selecting the transducers for the X...

Page 228: ...esultant field strength will be calculated and displayed Measurement of the resultant field strength in a transmission channel with large bandwidth To measure the resultant field strength in a transmi...

Page 229: ...are also possible that adapt the R S FSH to other transmission systems Operation Press the CHANNEL POWER key The R S FSH will open the measurement function menu To exit the channel power menu press t...

Page 230: ...e available the next time the USER standard is called When changing settings keep the following in mind The span is always coupled with the channel bandwidth If you change the bandwidth the R S FSH wi...

Page 231: ...or keys Confirm the selection with the ENTER key or the STANDARD softkey The R S FSH will open the window for entering the name of the USER standard Enter a name of your choice by using the numeric ke...

Page 232: ...wait Setting level for channel power measurement It will then set the optimum reference level Setting the channel bandwidth The channel bandwidth is used to define the bandwidth in which the R S FSH...

Page 233: ...the MANUAL SPAN softkey The largest span that is permitted in the channel power measurement is ten times the channel bandwidth At larger spans the result would be increasingly inaccurate because not...

Page 234: ...tivate the power display Press the POWER DISPLAY softkey Select PWR DISP ON by using the rotary knob or cursor keys and confirm with the POWER DISPLAY softkey or the ENTER key Unit for displaying the...

Page 235: ...r of data sets that can be stored by the number of data sets stored with calibration data Storage of calibration data can be selected in the SETUP menu see Chapter 2 section Saving Calibration Data If...

Page 236: ...aining free memory locations Free Locations are also displayed in the text box Since the data sets can be different in size the remaining storage space is indicated as a percentage value The data set...

Page 237: ...e data set and terminate the entry with the ENTER key The data set is saved to the R S FSH s internal memory under the name that has been given Loading measurement results Previously saved measurement...

Page 238: ...measurement mode print out the measurement and settings stored in the data set to a printer using PRINT press EXIT to quit the setting again When you press the EXIT softkey you return again to the di...

Page 239: ...y pressing the DELETE ALL DATASETS softkey Press the DELETE ALL DATASETS softkey Before all the data sets are deleted the R S FSH asks the user if he is sure that he wants to delete all the data sets...

Page 240: ...y The R S FSH opens the SAVE PRINT menu and the printout function offers to print out the current screen to a printer Instrument settings can also be saved and data sets loaded or deleted The SCREEN P...

Page 241: ...ith a certain on off ratio or a sawtooth However it is not at all obvious what the harmonic content of the signal is or if low level signals are superimposed This is easy to see with a spectrum analyz...

Page 242: ...hree mixing stages with the IFs 4031 MHz 831 25 MHz and 31 25 MHz Up to 3 GHz the R S FSH6 uses the same IFs as the R S FSH3 Between 3 GHz and 6 GHz it uses a first IF at 7231 MHz which it converts to...

Page 243: ...oser together than the bandwidth of the IF filter cannot be resolved This is why the bandwidth of the IF filter in a spectrum analyzer is referred to as the resolution bandwidth The R S FSH has resolu...

Page 244: ...is not affected The Figs below show a weak sine signal In the first Fig it is measured with a large video bandwidth and in the second with a narrow video bandwidth Limiting the video bandwidth smooth...

Page 245: ...are ignored On the basis of the operating principles of detectors a few recommendations can be made as to their use It is best to use the Auto Peak detector or the peak detector for spectrum analysis...