R&S NRP-Z5x
Remote Control - Commands
1137.0463.12
6.5
E-7
The query always yields
1
(=
OFF
).
Default setting
After a power-on reset, the zero offsets determined during the last calibration are used until the first
zeroing. Therefore, very slight zero offsets are to be expected with a warmed up sensor. Initialization by
means of
*RST
or
SYSTem:INITialize
has no influence on the current zero offsets.