Remote Control - Commands
R&S NRP-Z5x
1137.0463.12
6.22
E-7
Lower limit x and upper limit y of the trigger threshold
SENSe:CORRection:OFFSet:STATe
OFF
:
x = <lower test limit in W>
×
25
y = <upper test limit in W>
SENSe:CORRection:OFFSet:STATe
ON
:
x = <lower test limit in W>
×
25
×
10
OFFSET
/ 10 dB
y = <upper test limit in W>
×
10
OFFSET
/ 10 dB
<lower test limit in W>:
The nominal lower test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
<upper test limit in W>:
The nominal upper test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
The nominal test limits can be read out with the
SYSTem:INFO? "MINPOWER"
and
SYSTem:INFO?
"MAXPOWER"
command.
Default setting:
10
×
x
TRIGger:SLOPe[?] POSitive | NEGative
TRIGger:SLOPe
defines the edge of the trigger event with internal and external triggering. In this
connection, positive means increasing envelope power (with internal triggering) or increasing voltage
(with external triggering). This command has no effect in conjunction with trigger sources
BUS
,
HOLD
and
IMMediate.
The query yields
•
1
for
POSitive
,
•
2
for
NEGative
.
Default setting:
POSitive
TRIGger:SOURce[?] HOLD | IMMediate | INTernal | BUS | EXTernal |
TRIGger:SOURce
sets the trigger source.
•
HOLD
:
Triggering only with command
TRIGger:IMMediate
.
•
IMMediate
: Automatic triggering without explicit event.
•
INTernal
:
Triggering by the measurement signal. Relevant parameters:
TRIGger:LEVel,
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HYSTeresis,
TRIGger:HOLDoff
.
•
BUS
:
Triggering with command
*TRG
or
TRIGger:IMMediate
.
•
EXTernal
: Triggering via the hardware trigger bus, e.g. via the R&S NRP-Z3 USB adapter
or the R&S NRP base unit. Relevant parameters:
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HOLDoff
.
The query yields
•
1
for
HOLD,
•
2
for
IMMediate,
•
4
for
INTernal,
•
8
for
BUS,
•
16
for
EXTernal.
Default setting:
IMMediate