R&S NRP-Z5x
Remote Control - Commands
1137.0463.12
6.9
E-7
SENSe:AVERage:STATe[?] OFF | ON
SENSe:AVERage:STATe
switches on or off the averaging filter.
The query yields
•
1
for
OFF,
•
2
for
ON
.
Default setting:
ON
SENSe:AVERage:TCONtrol[?] MOVing | REPeat
SENSe:AVERage:TCONtrol
(terminal control)
defines the behaviour of the averaging filter. As soon as
a new measured value is shifted to the FIR filter, a new average value is available at the filter output,
which is obtained from the new measured value and the other values stored in the filter.
The
MOVing
parameter defines that each new average value is output as a measurement result. This
allows tendencies in the result to be recognized during the measurement procedure.
The
REPeat
parameter defines that a new result is output after the FIR filter has been filled with new
measured values. This ensures that no redundant information is output.
The query yields
•
1
for
MOVing
,
•
2
for
REPeat
.
Default setting:
REPeat
SENSe:CORRection:DCYCle[?] 0.001 to 99.999
SENSe:CORRection:DCYCle
sets the duty cycle to a percent value for the correction of pulse-
modulated signals. With the correction activated, the sensor calculates the signal pulse power from this
value and the mean power. Since the duty cycle is only useful in the
Continuous Average
mode, it is
evaluated only there.
The query yields the current duty cycle in percent.
Default setting:
1.0 [%]
SENSe:CORRection:DCYCle:STATe[?] OFF | ON
SENSe:CORRection:DCYCle:STATe ON
activates the duty cycle correction and thus the pulse-power
measurement whereas
SENSe:CORRection:DCYCle:STATe OFF
deactivates it.
The query yields
•
1
for
OFF,
•
2
for
ON
.
Default setting:
OFF