Description of the Menus
CTS60
1094.0006.60 6.40
E-5
6.7.4
Config Power Ramp Menu
In this menu the configuration for the power ramp measurement is defined. The template is shown at
the top of the screen. The various ranges of the template are identified with the letters B to F. The
tolerance values for these ranges are shown below the template. The limit B is defined as an absolute
level in
µ
W. The limits C to F are defined in dB and define levels relative to the normal transmit power.
Apart from the template the limits for the normal transmit power (NTP) of the DUT are also defined.
These are defined as absolute values in dBm and define the limits for NTP averaged over a burst.
The number of bursts (1 to 200) over which the NTP is averaged can also be configured in this menu.
The graphical display shows only one burst. The value set here is effective in the measurement menus
"Connection Established". "Power Ramp". and "RF Modulation".
Default
Press this key to reset all values to factory default values.
Menu up
Press this key to return to the previous menu.