CTS60
Remote Control/Annex C
1094.0006.60 C.23
E-5
Tolerance Range for Power-Time-Template
The level designations used in the power time template for the command
CALCulate:LIMit:POWer:TRANsmit[:TEMPlate][:DATA] can be obtained from the following illustration:
A
B
C
D
E
B
A
F
NTP
Syntax:
CALCulate:LIMit:POWer:TRANsmit[:TEMPlate][:DATA]
<numeric_value>, <numeric_value>,
<numeric_value>, <numeric_value>, <numeric_value>, <numeric_value>
Value range:
0.0 ... 100.0 nW
Absolute level at A
0.0 ... 100.0 uW
Absolute level at B
0.0 ... 10.0 dB
Relative level at C
0.0 ... 10.0 dB
Relative level at D
-10.0 ... 0.0 dB
Relative level at E
-20.0 ... 0.0 dB
Relative level at F
Default:
20.0 nW
25.0 uW
+4.0 dB
+1.0 dB
-1.0 dB
-6.0 dB
State:
ALL
Option:
none With query
Note
: The A-level is retained for compatibility reasons. The CTS60 doesn’t
support it. It is a dummy argument.