CTS60
Module Test
1094.0006.60
11.3
E-4
General Configuration information for Module Test can be displayed by pressing the “Config” softkey.
The following menu is then displayed.
Menu Conf. Module Test
Configuration information can be modified by selecting the desired field with the “Next Field” softkey
and editing its value.
The following fields are available for modification:
Trigger Mode
•
Free Run. The measurement is restarted again without any trigger condition being present.
•
Rising Slope. The measurement is started with rising power as the trigger event.
Trigger Delay
Setting of time values for the measurement delay after the trigger event from 0 to 10 ms in steps of 1 µs.
If the trigger mode is set to “Free Run”, then the value set for “Trigger Delay” is irrelevant. “Trigger
Delay” also allows displaying the rising edge of a periodical signal triggered with the condition “Rising
Slope”.
Meas. Window
Setting of discrete values between 25 µs and 1000 µs for the measurement window.
Exp Power
Setting of the maximum expected power between -10 dBm and 30 dBm in steps of 1 dB for fully utilising
the dynamic range of the test receiver of the CTS60.
Note:
If a value below the power of the measured signal is selected, the measured NTP value is not correct.
Mod.Ref
The reference frequency against which the modulation is measured can be selected. The values are:
•
Measured. The mean frequency of the measured signal. This mode is also used in manual test.
•
Nominal. The exact centre frequency of the carrier, as selected by the user.
•
Half Peak. The mid-point between the measured minimum and maximum measured peaks.
The default value is Measured.