MagNA Pure Compact Operator’s Manual - Version 1.3
12
A
A
1.2
Operating Environment
1.3
Operating Parameters
Note:
Exact values depend on the purification protocol used.
Temperatures allowed during
transportation/storage/packaging
-25°C to +70°C
Temperatures allowed during
operation
+18°C to +30°C
Pollution Degree: 2
Indoor use
Maximum relative humidity
80% (for operating temperatures up to 31°C);
decreases linearly to 50% for operating
temperatures up to 40°C
Altitude
up to 2000 m
Processing capability
1 to 8 samples per batch
Processing time
20 to 45 min (depending on protocol)
Pipetting Accuracy
25 µl to 100 µl:
ⱕ
5%
⬎
100 µl
ⱕ
2%
Sample Volume
100 – 1000 µl
Elution Volume
50 – 200 µl
Internal Control Volume
5 – 20 µl
Specifications of the Instrument
Operating Environment
MagNA Pure Compact Manual.book Seite 12 Dienstag, 28. Februar 2006 4:16 16