Maintenance and Trouble Shooting
81
C
C
2.
Service Maintenance
We recommend that a Roche Diagnostics service engineer perform the first Service Main-
tenance of the instrument one year after the instrument is installed. Subsequently, this
Service Maintenance should be performed annually.
Note:
Service Maintenance may only be performed by a Roche Diagnostics service engi-
neer. Contact your local Roche representant for maintenance assistance and for more
details on country-specific service contracts.
During Service Maintenance, the Roche service engineer will do the following:
1. Check and (if necessary) adjust:
䉴
Easy Teaching procedure
䉴
Clot sensor signals
䉴
Temperature settings on Heating Unit
䉴
Leakage test
䉴
Tip alignement
䉴
Belt tension
2. Exchange:
䉴
O-Rings
䉴
Hepa filter
䉴
UV lamp
䉴
Air filter
3. Clean and/or grease:
䉴
Clot sensor
䉴
LM Guide
䉴
Z-axis thread bar/slide shafts
䉴
Z-plunger
Service Maintenance